1 October 2011 Mueller-matrix microscopic system based on electro-optics modulation for cellular measurement
Yubo Li, Fengyuan Zou, Huaxing Jiang, Qiang Zhou, Xiaoqing Jiang, Jianyi Yang
Author Affiliations +
Abstract
We propose a design of a Mueller-matrix-microscopic system that can efficiently capture the parameter image with two liquid-crystal modulators (LCMs). The usage of the LCMs can significantly reduce the measurement time for Mueller-matrix parameters. This system employs one LCM to realize real-time measurement of all subimages derived from the Stokes vector and the other one to change the light sources. Seven experimental results, acquired by using incident light with LCM-modulated light sources, are enough for biological research. An error analysis for m44 of the Mueller matrix is involved in order to present the error evaluation. The corresponding error evaluation shows that the measurement errors could be controlled under 0.07 deg.
©(2011) Society of Photo-Optical Instrumentation Engineers (SPIE)
Yubo Li, Fengyuan Zou, Huaxing Jiang, Qiang Zhou, Xiaoqing Jiang, and Jianyi Yang "Mueller-matrix microscopic system based on electro-optics modulation for cellular measurement," Optical Engineering 50(10), 103202 (1 October 2011). https://doi.org/10.1117/1.3638060
Published: 1 October 2011
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Modulation

Electro optics

Error analysis

Modulators

Light sources

Polarization