1.INTRODUCTIONESA S mission CHEOPS1, aims to improve our knowledge, of exoplanet radii through the high-precision photometric monitoring of their transits. This will allow to better constrain the density of these planets and help to better distinguish between rocky planets, gaseous giants and all the possible intermediate zoo of planets. The satellite has access to a large fraction of the sky, thus it will be able to characterize a large part of the transiting planets, already known and to be discovered by the next exoplanet discovering machines such as TESS2 and PLATO3, or ground based detection surveys such a NGTS4, SPECULOOS5 and others. CHEOPS payload is a simple defocused 32-cm telescope. In order to fulfill the ambitious science goals of the mission, it has to reach an absolute photometric precision of 20 ppm. The focal plane is a CCD-47-20 from E2V. In order to ensure this performance level, it is necessary to maintain the detector temperature to a stability level around 10 mK, and the detector bias voltage stability to a few microvolts. The pointing of the instrument is done with reaction wheels. Its stability cannot be better than a few arcsecond. As a consequence, the PSF of the instrument will move on the detector. It is thus necessary to measure the Flat field of the instrument with 0.1% precision in order to be able to recover the 20-ppm precision required. Only a 200x200 portion of this array will be transmitted to the ground, however we have performed calibration on the full area of the CCD as the possibility to move the scientific window is a part of the instrument redundancy. It will in particular allow to avoid hot pixels that will appear in orbit. After a successful CCD test6 campaign, a test bench7 to perform the needed instrument calibration and to test the instrument performances has been built. We report here the results of this calibration campaign. We could perform the following measurements and characterizations: 2.BIAS – DARK – LINEARITY MEASUREMENTSThe calibrations campaign took place at the beginning of 2018. It has been the first time the full instrument was assembled. The detector chain has been characterized with the real flight electronics. The measurements were performed in a similar fashion as in our CCD tests6. Results are shown in the Table 1. Two reading speeds are implemented, 230 kHz for bright stars and 100 kHz for faint stars. Table 1Summary of the measured characteristics of the CHEOPS instrument. Two reading speeds are available on board, a fast mode for bright stars at 230 kHz and a slow reading mode for the faint stars at 100 kHz.
The way the CCD saturate has been a surprise. During the CCD tests6, the full well of the CCD was reached in average at the same level all across the CCD. With the flight electronics, the situation has dramatically changed, the full well seem to be reached sooner in the center of the CCD. Blooming starts in the center before than on the borders. In addition, before the saturation is reached, a fix pattern noise appears, lowering a bit more the available dynamic. The origin of this fix pattern noise is unknown. It had already been seen on ULTRACAM8 with the same E2V CCD and called peppering effect. Figure 1 shows the map of the usable dynamic on the CCD. The consequence of this effect, and the fact that the reading electronics gain is slightly too high, makes it very difficult to do photometry with saturated stars, which would have been a desirable possibility of the system. The gain is identical at both reading speeds, and as far we could test the same in science window mode. The linearity for the fast reading mode is shown on Figure 2. 3.FLAT-FIELDSOne of the most important calibration to perform was the flat fielding of the instrument. As the stellar spectrums vary substantially we set out to build high-precision color-dependent flat-fields measurements. For this we used an integrating sphere at the focus of the calibration set-up7. To have a final precision of 0.1% many steps were necessary. While this integrating sphere had a diameter of 280 mm, and a port of only 20 mm, we could measure a radial non-uniformity of its illumination. We measured this using a monochromator and a dedicated scanning set-up to make a 2D map of the sphere illumination, at the required precision. We could model non-uniformity with a radial function (see Figure 3). Doing successfully this measurement, that lasted a week, and having consistent result proved the reliability of our set-up. In order to be able to apply this correction to the images we used a mechanical reference so that it was possible to calculate the correction in the pixel space. The optics of the telescope have a non-negligible level of distortion. This has photometric consequences on the flat-field measured. There is an excess of light on the border of the image, because of barrel distortion. It is necessary to characterize properly the distortion because when doing aperture photometry with a star (a point source), the flat-field need to be debiased as the detector is not actually more sensitive on the border. Distortion has been measured using a standard high precision grid of point stamped on a glass plate. It has been installed in front of the integrating sphere. With this measurement it was possible to determine the optical center of the field and a second order model of the distortion. The pixel response non-uniformity of a CCD varies significantly with wavelength6. To improve the final precision of the flat-field corrections we made a color dependent measurement. We measured flat-fields using a mix of standard photometric filters (Johnson UBVRI filters) and monochromator defined wavelengths (there are 43 wavelengths from 440 nm to 950 nm). It is thus possible to reproduce smooth stellar spectrums. It has also been necessary to calibrate the spectrum of the white source that was used, the transmission of the instrument and the Quantum Efficiency of the detector. A particular care has been taken to understand the fringing of the detector. Using the monochromator at a high sampling in wavelength it has been possible to determine the frequency of the wavelength periodicity of the fringe contrast inversion. This allows to estimate the systematic error one does in sampling the flat field in the fringing region. We could find a periodicity of 4.55 nm. Thus, in order to limit the systematic errors due to fringing it is necessary to integrate over the full I band (see Figure 5). In order to prove that we could recombine the flat field we could make a measurement using a tungsten lamp as a source. Once its spectrum calibrated it is possible to determine the combination of flat-fields necessary to approximate at best the lamp spectrum (see Figure 5). If we compare the measured flat-field (Figure 6) and the recombined one the error is well below 0.1% and very near to the photonic limit. 4.PSF MEASUREMENTSPSF measurements were performed using the test bench in point source mode. It allows to check that the optical alignment of the telescope is in line with expectations. The measurements suffer from some issues: the pupil uniformity of the test bench was suboptimal (12% center to border), the optical quality of the bench was by design around λ/3, and the telescope was pointing horizontally in the test chamber, thus the gravity affect the aberrations of the PSF. Only in-flight measured PSF could be used to do any kind of scientific data analysis. In the center of the field, the PSF encircled energy radius at 90% is 12.8 ±0.3 pixels at 680 nm. The PSF, in one exposure, can hold ~ 4.5 106 e- (this is an order of magnitude). The low-flux large-scale diffraction halo that can be seen on the Figure 7 is probably due to the mid-spatial frequency wavefront errors that are present on the primary mirror and will probably remain in space. 5.PHOTOMETRIC STABILITYWe had set-up7 to prove experimentally the end-to-end photometric stability of the instrument. In order to perform such a measurement, we have developed a Super Stable Source9 (SSS), a closed loop system to stabilize the source of the test bench to the ppm level. The full test would have been to illuminate the instrument with a star like source, and thanks to precise control of the field and the pupil alignment of the test bench, the test bench induced photometric variations would have remained small enough to show the instrument performances. The plan was also to perform the test adding jitter to the PSF position, and slowly changing the temperature of the telescope to mimic the in-orbit environment. Because of time constraints we had to limit the number of tests and the full end-to-end measurement was not performed. It has been however possible to make some long-term stability measurement using an integrating sphere as a source. Because of issues that had not been understood at the moment of the measurement we used the SSS in open loop, but we recorded the photometric control signal as well as some environmental data. The instrument house-keeping data were also recorded. It is thus possible to detrend the data acquired with the instrument from most of the test bench induced photometric variations. The first approach to check the stability of the instrument is to check the stability of the CCD temperature, and its reading electronics. It is also necessary to check the stability of the bias voltages. The Table 2 below summarize the performances and the requirements. Everything is well in the requirement. Table 2Stability of the different electrical and thermal parameter of the instrument detector.
On Figure 8 one can see the nearly exact correspondence between the photometric data from the instrument and the one from the reference photometer. However, once the ratio of the two measurements done, there is still a low frequency structure. This structure is correlated with the temperature in the lab. This can be explained by an unfortunate property of the optical fiber that was used during the measurement. We did not realize when we chose the fiber that it had a double cladding. It means that the fiber has a double guiding structure and that light is transported not only in the core but also in the first cladding. More over the second guiding structure has an aperture that is twice as big as the one of the core guiding structure. The light used for the reference detector is directly picked up inside the integrating sphere. The second guiding structure, an interface between silica glass and plastic, has its aperture that is very sensitive to temperature, and thus the flux going to the reference photometer increases with temperature (and the ratio shown on Figure 8 decreases). Once established the correlation, one can correct for it and analyze the resulting photometry. Figure 9 shows the noise properties of the resulting photometry. A noise floor of 25 ppm noise in 6h is reached, this is slightly above the requirement, but the remaining structure in the data seem still corelated to the different temperatures in the lab. Further data reduction might improve this result. In any case this is a clear sign that the limiting factor in the measurement is the test bench and not the instrument. We had no time to change the fiber and perform and other tests. 6.CONCLUSIONWe have presented here the results of CHEOPS calibration campaign. We have measured the performances of the detector chain. We have shown in particular that its behavior near to saturation will make it difficult to do saturated star photometry. The flat-field with a better than 0.1% precision and with the current data it is possible to produce flat-fields adapted to the color of the observed star. We have demonstrated that the instrument intrinsic precision is better than 25 ppm. REFERENCESBroeg, C., Fortier, A., Ehrenreich, D., Alibert, Y., Baumjohann, W., Benz, W., Deleuil, M., Gillon, M., Ivanov, A., Liseau, R., Meyer, M., Oloffson, G., Pagano, I., Piotto, G., Pollacco, D., Queloz, D., Ragazzoni, R., Renotte, E., Steller, M., et al.,
“CHEOPS: A transit photometry mission for ESA’s small mission programme,”
in European Physical Journal Web of Conferences,
03005
(2013). Google Scholar
Ricker, G. R., Winn, J. N., Vanderspek, R., Latham, D. W., Bakos, G. Á., Bean, J. L., Berta-Thompson, Z. K., Brown, T. M., Buchhave, L., Butler, N. R., Butler, R. P., Chaplin, W. J., Charbonneau, D., Christensen-Dalsgaard, J., Clampin, M., Deming, D., Doty, J., De Lee, N., Dressing, C., et al.,
“Transiting Exoplanet Survey Satellite (TESS),”
Space Telescopes and Instrumentation 2014: Optical, Infrared, and Millimeter Wave, 9143
(914320),
(2014). Google Scholar
Rauer, H., Catala, C., Aerts, C., Appourchaux, T., Benz, W., Brandeker, A., Christensen-Dalsgaard, J., Deleuil, M., Gizon, L., Goupil, M.-J., Güdel, M., Janot-Pacheco, E., Mas-Hesse, M., Pagano, I., Piotto, G., Pollacco, D., Santos, C., Smith, A., Suárez, J.-C., et al.,
“The PLATO 2.0 mission,”
Experimental Astronomy, 38 249
–330
(2014). https://doi.org/10.1007/s10686-014-9383-4 Google Scholar
Wheatley, P. J., West, R. G., Goad, M. R., Jenkins, J. S., Pollacco, D. L., Queloz, D., Rauer, H., Udry, S., Watson, C. A., Chazelas, B., Eigmüller, P., Lambert, G., Genolet, L., McCormac, J., Walker, S., Armstrong, D. J., Bayliss, D., Bento, J., Bouchy, F., et al.,
“The Next Generation Transit Survey (NGTS),”
mnras, 475 4476
–4493
(2018). https://doi.org/10.1093/mnras/stx2836 Google Scholar
Gillon, M., Jehin, E., Delrez, L., Magain, P., Opitom, C. and Sohy, S.,
“SPECULOOS: Search for habitable Planets EClipsing ULtra-cOOl Stars,”
in presented at Protostars and Planets VI Posters,
2
(2013). Google Scholar
Deline, A., Sordet, M., Wildi, F. and Chazelas, B.,
“The testing and characterization of the CHEOPS CCDs,”
in Society of Photo-Optical Instrumentation Engineers (SPIE) Conference Series,
(2017). Google Scholar
Wildi, F. P., Chazelas, B., Deline, A., Sordet, M. and Sarajlic, M.,
“The CHEOPS instrument on-ground calibration system,”
Techniques and Instrumentation for Detection of Exoplanets VII, 9605
(96051B),
(2015). Google Scholar
Feline, W. J.,
“ULTRACAM photometry of eclipsing cataclysmic variable stars,”
(2008). Google Scholar
Wildi, F. P., Deline, A. and Chazelas, B.,
“A white super-stable source for the metrology of astronomical photometers,”
Techniques and Instrumentation for Detection of Exoplanets VII, 9605
(96051T),
(2015). Google Scholar
|