Paper
22 January 1993 Spectral radiance source for vacuum ultraviolet calibrations
Fred G. Sherrell
Author Affiliations +
Abstract
A working standard extended source of spectral radiance has been assembled and calibrated for use in the 150- to 260-nm region. The assembly consists of a 10-kw, argon arc lamp mounted to irradiate a 4-in. diameter diffuser. The reflected energy from the diffuser is utilized as a calibration source. The procedure for calibrating the spectral radiance of the diffuser is reviewed. Traceability is accomplished by the use of a standard vacuum photodiode. The total uncertainty in the calibration of the source is estimated to be +/- 15 percent.
© (1993) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Fred G. Sherrell "Spectral radiance source for vacuum ultraviolet calibrations", Proc. SPIE 1764, Ultraviolet Technology IV, (22 January 1993); https://doi.org/10.1117/12.140855