Paper
31 December 1980 Optical Alignment Of High Resolution Fourier Transform Spectrometers
J. B. Breckinridge, F. G. O'Callaghan, A. G. Cassie
Author Affiliations +
Abstract
Remote sensing, high resolution FTS instruments often contain three primary optical sub-systems: Fore-Optics, Interferometer Optics, and Post, or Detector Optics. We discuss the alignment of a double-pass FTS containing a cat's-eye retroreflector. Also, the alignment of fore-optics containing confocal paraboloids with a reflecting field stop which relays a field image onto a camera is discussed.
© (1980) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. B. Breckinridge, F. G. O'Callaghan, and A. G. Cassie "Optical Alignment Of High Resolution Fourier Transform Spectrometers", Proc. SPIE 0251, Optical Alignment I, (31 December 1980); https://doi.org/10.1117/12.959456