26 September 1995 Use of INM cryogenic radiometer to calibrate transfer standard detectors at 1550 nm
Oualid Touayar, J. M. Coutin, M. Jean Bastie
Author Affiliations +
Abstract
At first, this paper gives a brief description of the cryogenic radiometer used as standard for radiometric measurements at the Institut National de Metrologie (INM). After, it describes the procedure used to calibrate transfer detectors, for the measurements in the field of optical fiber telecommunications. This absolute calibration was performed at the laser diode wavelength of 1550 nm. The first results for three different detectors calibrated by comparison to INM cryogenic radiometer for the above wavelength shows that the uncertainty of the calibration is less than 0.5% at 1 (sigma) level and is about twice as good as the uncertainty given by the former extrapolation method used at INM.
© (1995) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Oualid Touayar, J. M. Coutin, and M. Jean Bastie "Use of INM cryogenic radiometer to calibrate transfer standard detectors at 1550 nm", Proc. SPIE 2550, Photodetectors and Power Meters II, (26 September 1995); https://doi.org/10.1117/12.221419