Paper
28 June 2001 Development of an automated method for evaluation of sharpness of digital radiographs using edge method
Author Affiliations +
Abstract
The presampling modulation transfer function (MTF) can be determined by the edge spread function in which the sampling interval is narrower than the pixel-to-pixel interval from slight angled edge image. It is important that the precision of the presampling MTF depend on the precision of the edge angle. In this study, we have developed the automated method, which includes a precise edge angle determination process for the measurement of the presampling MTF.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ikuo Kawashita, Kouji Maeda, Hidetaka Arimura, Kaoru Morikawa, and Takayuki Ishida "Development of an automated method for evaluation of sharpness of digital radiographs using edge method", Proc. SPIE 4320, Medical Imaging 2001: Physics of Medical Imaging, (28 June 2001); https://doi.org/10.1117/12.430929