Paper
10 February 2006 UV radiation dose measurements
R. Cardoso, J. P. Valdez-Chaparro, E. Rosas
Author Affiliations +
Proceedings Volume 6046, Fifth Symposium Optics in Industry; 60461R (2006) https://doi.org/10.1117/12.674596
Event: Fifth Symposium Optics in Industry, 2005, Santiago De Queretaro, Mexico
Abstract
We present the implementation of the ultraviolet (UV) radiation dose measurement system at the Laboratorio de Dosimetria Optica (LDOS) of the Centro Nacional de Metrologia (CENAM). The UV dose measurement ranges established so far cover from 12 Jm-2 to 7 256 Jm-2 and from 171 Jm-2 to 102 842 Jm-2, for 254 nm and 365 nm, respectively; corresponding to two emission lines of an arc Hg(Xe) lamp. Those dose ranges can be reached with exposure times from 2 s to 10 min and the reference dose uncertainty has been estimated to be of ±5,4 %.
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
R. Cardoso, J. P. Valdez-Chaparro, and E. Rosas "UV radiation dose measurements", Proc. SPIE 6046, Fifth Symposium Optics in Industry, 60461R (10 February 2006); https://doi.org/10.1117/12.674596
Lens.org Logo
CITATIONS
Cited by 6 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace