Paper
19 May 2006 InGaAs trap detector used as a near-infrared transfer standard for detector calibrations
Ji Wang, Shuang Li, Lei Zhang, Feng Chen, Xiaobing Zheng
Author Affiliations +
Abstract
This paper presents results of investigations of the radiant power spectral responsivity within the near-infrared spectral range. Large-area windowless InGaAs photodiodes of 10mm×10mm are studied in the spectrum ranged from 900 nm to 1600 nm. The InGaAs trap detector used as a transfer standard for power measurements has been characterized and calibrated, obtaining an uncertainty of less than 0.1% at several wavelengths.
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ji Wang, Shuang Li, Lei Zhang, Feng Chen, and Xiaobing Zheng "InGaAs trap detector used as a near-infrared transfer standard for detector calibrations", Proc. SPIE 6150, 2nd International Symposium on Advanced Optical Manufacturing and Testing Technologies: Optical Test and Measurement Technology and Equipment, 61501D (19 May 2006); https://doi.org/10.1117/12.676739
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Indium gallium arsenide

Radiometry

Photodiodes

Calibration

Cryogenics

Infrared radiation

Back to Top