Paper
6 August 2009 The performance test and the application of CCD
Shu-rong Wang, Xiao-hu Yang
Author Affiliations +
Abstract
The e2v technologies' CCD sensor CCD57-10, which will be used as a detector in the Limb UV Imaging Spectrometer, is tested at first. Then the spectral responsivity calibration of the whole Imaging Spectrometer simply is described. The error analysis and the correction of the test of CCD's parameters under good repeatability and stability indicate that the properties of the CCD are suitable to the Limb UV Imaging Spectrometer. The spectral responsivity calibration and the theoretical calculation of the whole Imaging Spectrometer are within an acceptable error level.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Shu-rong Wang and Xiao-hu Yang "The performance test and the application of CCD", Proc. SPIE 7384, International Symposium on Photoelectronic Detection and Imaging 2009: Advances in Imaging Detectors and Applications, 73841M (6 August 2009); https://doi.org/10.1117/12.835592