Paper
28 December 2010 Application of Monte Carlo method for analyzing reference involute's measurement accuracy
Zhifeng Lou, Liding Wang
Author Affiliations +
Proceedings Volume 7544, Sixth International Symposium on Precision Engineering Measurements and Instrumentation; 75446N (2010) https://doi.org/10.1117/12.885376
Event: Sixth International Symposium on Precision Engineering Measurements and Instrumentation, 2010, Hangzhou, China
Abstract
In all accuracy indexes of gears, measurement of involute tooth profile's error is a difficult technical problem. A doubledisc instrument for measuring involute is introduced, which can be used to measure reference involute. The main errors sources of the instrument are analyzed, and all measurement errors are simulated by Mont-Carlo method. After analyzing error sequence, the double disc instrument's measurement uncertainty is 0.45μm when measuring gear (m=4, z=30, α=20°), which is less than the value evaluated by GUM method, and more reliable and accurate.
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Zhifeng Lou and Liding Wang "Application of Monte Carlo method for analyzing reference involute's measurement accuracy", Proc. SPIE 7544, Sixth International Symposium on Precision Engineering Measurements and Instrumentation, 75446N (28 December 2010); https://doi.org/10.1117/12.885376
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Error analysis

Teeth

Monte Carlo methods

Laser applications

Device simulation

Instrument modeling

Laser development

Back to Top