Paper
28 November 2011 Research on coupling efficiency of optical coherence tomography system with gradient refractive index rod lens
Zhihui Xu, Hao Lv
Author Affiliations +
Abstract
Coupling efficiency equation of optical coherence tomography system with gradient refractive index rod lens is established and discussed theoretically and experimentally. The experimental result is consistent with the theoretical result. The higher coupling efficiency of optical coherence tomography system can be obtained if the θ which is the angle between mirror and y axis should be smaller than 0.5 degree.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Zhihui Xu and Hao Lv "Research on coupling efficiency of optical coherence tomography system with gradient refractive index rod lens", Proc. SPIE 8202, 2011 International Conference on Optical Instruments and Technology: Solid State Lighting and Display Technologies, Holography, Speckle Pattern Interferometry, and Micro/Nano Manufacturing and Metrology, 820219 (28 November 2011); https://doi.org/10.1117/12.911959
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical coherence tomography

GRIN lenses

Mirrors

Refractive index

Gaussian beams

Single mode fibers

Fiber couplers

Back to Top