1 May 1996 Checking-up of optical graduated rules by laser interferometry
Author Affiliations +
The main aspects related to the operating principle, design, and implementation of high-productivity equipment for checking-up the graduation accuracy of optical graduated rules used as a length reference in optical measuring instruments for precision machine tools are presented. The graduation error checking-up is done with a Michelson interferometer as a length transducer. The instrument operation is managed by a computer, which controls the equipment, data acquisition, and processing. The evaluation is performed for rule lengths from 100 to 3000 mm, with a checking-up error < 2 ?m/m. The checking-up time is about 15 min for a 1000-mm rule, with averaging over four measurements.
Nicolae Miron and Dan G. Sporea "Checking-up of optical graduated rules by laser interferometry," Optical Engineering 35(5), (1 May 1996). https://doi.org/10.1117/1.600632
Published: 1 May 1996
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Laser optics

Microscopes

Data acquisition

Human-machine interfaces

Transducers

Laser interferometry

Photodetectors

RELATED CONTENT


Back to Top