4 December 2020 Comparative analysis of measurement standards among laboratories for combustible gas detection alarms
Yao Yao, Zhipeng Wang, Yuyu Tian
Author Affiliations +
Proceedings Volume 11617, International Conference on Optoelectronic and Microelectronic Technology and Application; 1161741 (2020) https://doi.org/10.1117/12.2587267
Event: International Conference on Optoelectronic and Microelectronic Technology and Application, 2020, Nanjing, China
Abstract
This article, based on JJG 693-2011 Verification Regulation of Combustible Gas Detection Alarms, is to design the measurement comparison scheme, measuring the same blind sample of combustible gas detection alarm by measurement standards from different laboratories, make statistic analysis of experimental data by Z value, and obtain the comparison results, which provides an effective guarantee for the accuracy of combustible gas detection alarms and the traceable transmitting of measuring instruments in the future.
© (2020) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yao Yao, Zhipeng Wang, and Yuyu Tian "Comparative analysis of measurement standards among laboratories for combustible gas detection alarms", Proc. SPIE 11617, International Conference on Optoelectronic and Microelectronic Technology and Application, 1161741 (4 December 2020); https://doi.org/10.1117/12.2587267