Paper
18 April 2021 X-ray interferometry technique for thin-films study using an x-ray microfocus laboratory source
Author Affiliations +
Abstract
We are proposing an X-ray reflecto-interferometry technique using an x-ray microfocus laboratory source for thin-film structure research based on compound refractive lenses. The idea of this technique is to use a very simplified experimental setup where a focused X-ray beam reflected from parallel flat surfaces creates an interference pattern in a wide angular range. Due to this, the interference pattern can be obtained in a single shot without the need to rotate the specimen or the detector. The applicability of this technique has been demonstrated using the MetalJet Excillium microfocus laboratory source, which has GaKα emission line at 9.25 keV. A series of interference patterns for Si3N4 membranes and the experimentally obtained film thickness are in good agreement with the declared characteristics. The main advantages and future possible of the reflecto-interferometry technique are discussed.
© (2021) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
M. Voevodina, S. Lyatun, D. Zverev, A. Barannikov, I. Lyatun, I. Snigireva, and A. Snigirev "X-ray interferometry technique for thin-films study using an x-ray microfocus laboratory source", Proc. SPIE 11776, EUV and X-ray Optics, Sources, and Instrumentation, 1177609 (18 April 2021); https://doi.org/10.1117/12.2589495
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

X-ray sources

Thin films

Interferometry

X-ray optics

Sensors

Multilayers