Paper
6 December 2022 Studies on calibration of infant phototherapy incubator
Author Affiliations +
Proceedings Volume 12458, International Conference on Biomedical and Intelligent Systems (IC-BIS 2022); 1245826 (2022) https://doi.org/10.1117/12.2660324
Event: International Conference on Biomedical and Intelligent Systems, 2022, Chengdu, China
Abstract
Infant phototherapy incubator is widely used in medical institutions to reduce the concentration of bilirubin in the body of infant while maintaining the body temperature of the infant. This article presents a novel design of calibration procedure to measure the instruments’ metrological parameters such as "error of total irradiance for bilirubin after pre-ageing", "relative local distribution of total irradiance for bilirubin after pre-ageing", "uniformity of incubator temperature", "accuracy of skin temperature sensor", "relative humidity indication error" and "accuracy of oxygen sensor" in order to ensure the accuracy and reliability of infant phototherapy incubator. The experimental results show that the calibration procedure presented in this article can ensure the metrological traceability of infant phototherapy incubators in daily use.
© (2022) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Tingting Ren, Yang Xu, Han Luo, Delong Hu, and Feng Chen "Studies on calibration of infant phototherapy incubator", Proc. SPIE 12458, International Conference on Biomedical and Intelligent Systems (IC-BIS 2022), 1245826 (6 December 2022); https://doi.org/10.1117/12.2660324
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Phototherapy

Calibration

Temperature sensors

Sensors

Metrology

Back to Top