Paper
18 December 1995 Coherence effects in modeling laser Doppler perfusion flowmetry
Author Affiliations +
Abstract
When performing direct-contact laser-Doppler flowmetry on experimental flow models, the power spectra of the detector signal can be obtained by homodyne or by heterodyne detection. Especially with uniformly moving probe particles coherence effects are observed, leading to changes in the width of the power spectrum. Due to destructive interference effects at the detector surface, in the measured homodyne spectra the contribution of relatively high Doppler frequencies is suppressed compared with that of lower frequencies. This results in spectra which are narrower than expected theoretically. This effect allows us to investigate the importance of the relative amount of coherence areas at the surface of the detector.
© (1995) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Frits F. M. de Mul, Wiendelt Steenbergen, T. Vonck, and Jan Greve "Coherence effects in modeling laser Doppler perfusion flowmetry", Proc. SPIE 2626, Photon Propagation in Tissues, (18 December 1995); https://doi.org/10.1117/12.228677
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Photons

Doppler effect

Homodyne detection

Heterodyning

Scattering

Sensors

Monte Carlo methods

Back to Top