11 April 1996 Characterization of a C-arm-mounted XRII for 3D image reconstruction during interventional neuroradiology
Author Affiliations +
Abstract
Treatment of subarachnoid aneurysms with endovascular techniques (e.g. placement of Guglielmi coils) is currently limited by the inability to visualize the neck of the aneurysm after the initial coils have been placed. Coils projecting into parent vessels may cause thrombosis, while incomplete filling leads to regrowth of the aneurysm. Since the procedure is performed using a gantry-mounted x-ray image intensifier (XRII), we have used this system to obtain 2- dimensional (2-D) images over approximately 200 degrees and reconstructed a 3-D image of the embolizing coil, residual aneurysm, and the parent vessel. The required data can be acquired, reconstructed and presented to the neuroradiologist during the interventional procedure. We have characterized an existing clinical C-arm radiographic system, and have developed correction procedures which provide a consistent and adequate data set for standard CT reconstruction using convolution-backprojection techniques. The angle of acquisition for each image is recorded using a hub-mounted angle encoder accurate to within plus or minus 0.3 degrees. We have characterized the XRII distortion over the rotation, and can correct images acquired at any known angle to within plus or minus 0.06 pixels. We have also measured the motion of the center of rotation (reproducible to within plus or minus 0.13 pixels) and correct for the displacement using an image shift and interpolation algorithm. Finally, we have investigated the effects on CT reconstruction of variable dilutions of contrast agent during the cardiac cycle, and have shown the contrast-to-noise ratio (CNR), in the absence of photon noise, to be 28. This is larger than the CNR which we achieve due to photon noise alone.
© (1996) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Rebecca Fahrig, Allan J. Fox, and David W. Holdsworth "Characterization of a C-arm-mounted XRII for 3D image reconstruction during interventional neuroradiology", Proc. SPIE 2708, Medical Imaging 1996: Physics of Medical Imaging, (11 April 1996); https://doi.org/10.1117/12.237797
Lens.org Logo
CITATIONS
Cited by 11 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Distortion

3D image reconstruction

Computer programming

Image acquisition

Data acquisition

X-rays

Sensors

Back to Top