3 May 2000 Optical measurement of photosensitizer concentration using a probe with a small source-detector fiber separation
Author Affiliations +
Abstract
We present a novel minimally invasive method to measure the concentration of photodynamic therapy drugs in real time. The method is based on measurements of backscattered and fluorescence light using a steady state fluorescence spectrometer. The ratio of the fluorescence to scattered light is found to be linearly proportional to the absorption coefficient of the photosensitizer. The fiber-optic probe used for the measurements has a small source-detector separation, therefore the measurements could be performed through the working channel of an endoscope.
© (2000) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Murat Canpolat and Judith R. Mourant "Optical measurement of photosensitizer concentration using a probe with a small source-detector fiber separation", Proc. SPIE 3911, Biomedical Diagnostic, Guidance, and Surgical-Assist Systems II, (3 May 2000); https://doi.org/10.1117/12.384915