14 June 2002 Low-sidelobe limited diffraction optical coherence tomography
Author Affiliations +
Abstract
Optical Coherence Tomography (OCT) is a relatively new type of imaging system for medical diagnosis. Because most current OCT systems use a sharply focused beam in tissues, they have a short depth of field (high image resolution is near the focus only). In this paper, limited diffraction beams of different orders are used to increase depth of field and to reduce sidelobes in OCT. Results show that the proposed OCT system has a lateral resolution of about 4.4 wavelengths (the central wavelength of the source is about 940 nm with a bandwidth of about 70 nm) and lower than -60 dB sidelobes over an entire depth of field of 4.5 mm with the diameter of the objective lens of 1 mm.
© (2002) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jian-yu Lu, Jiqi Cheng, Yanfang Li, and Brent D. Cameron "Low-sidelobe limited diffraction optical coherence tomography", Proc. SPIE 4619, Coherence Domain Optical Methods in Biomedical Science and Clinical Applications VI, (14 June 2002); https://doi.org/10.1117/12.470498
Lens.org Logo
CITATIONS
Cited by 12 scholarly publications and 1 patent.