Paper
29 July 2004 Laser Doppler blood flowmetry with FPGA processing
C. Kongsavatsak, D. He, Stephen P. Morgan, Barrie R. Hayes-Gill, John A. Crowe, Matt Clark
Author Affiliations +
Proceedings Volume 5486, ALT'03 International Conference on Advanced Laser Technologies: Biomedical Optics; (2004) https://doi.org/10.1117/12.571996
Event: ALT'03 International Conference on Advanced Laser Technologies: Biomedical Optics, 2003, Silsoe, United Kingdom
Abstract
A single channel laser Doppler blood flowmetry device has been implemented using a photodetector linked to a field programmable gate array. Filters (low pass, band pass and frequency weighted) have been developed for processing Doppler signals to obtain flow and concentration measurements. The responses of these filters are demonstrated using measurements from modulated light signals, a rotating diffusing disc and in vivo measurements of blood flow.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
C. Kongsavatsak, D. He, Stephen P. Morgan, Barrie R. Hayes-Gill, John A. Crowe, and Matt Clark "Laser Doppler blood flowmetry with FPGA processing", Proc. SPIE 5486, ALT'03 International Conference on Advanced Laser Technologies: Biomedical Optics, (29 July 2004); https://doi.org/10.1117/12.571996