13 March 2009 2D array transducers for real-time 3D ultrasound guidance of interventional devices
Edward D. Light, Stephen W. Smith
Author Affiliations +
Abstract
We describe catheter ring arrays for real-time 3D ultrasound guidance of devices such as vascular grafts, heart valves and vena cava filters. We have constructed several prototypes operating at 5 MHz and consisting of 54 elements using the W.L. Gore & Associates, Inc. micro-miniature ribbon cables. We have recently constructed a new transducer using a braided wiring technology from Precision Interconnect. This transducer consists of 54 elements at 4.8 MHz with pitch of 0.20 mm and typical -6 dB bandwidth of 22%. In all cases, the transducer and wiring assembly were integrated with an 11 French catheter of a Cook Medical deployment device for vena cava filters. Preliminary in vivo and in vitro testing is ongoing including simultaneous 3D ultrasound and x-ray fluoroscopy.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Edward D. Light and Stephen W. Smith "2D array transducers for real-time 3D ultrasound guidance of interventional devices", Proc. SPIE 7265, Medical Imaging 2009: Ultrasonic Imaging and Signal Processing, 72650N (13 March 2009); https://doi.org/10.1117/12.808225