15 June 1988 Technicon H*1 Hematology System: Optical Design Considerations
G M Colella, D H Tycko, W Groner
Author Affiliations +
Proceedings Volume 0907, Laser Surgery: Characterization and Therapeutics; (1988) https://doi.org/10.1117/12.945330
Event: 1988 Los Angeles Symposium: O-E/LASE '88, 1988, Los Angeles, CA, United States
Abstract
The Technicon H*1 systemTM is a clinical laboratory flow cytometer which performs a complete hematology profile, providing quantitative information on the various types of cells in a blood sample. A light-scattering method, using a HeNe laser, determines in a single flow channel the red cell count, platelet count, and the distributions of red cell volume, red cell hemoglobin concentration, and platelet volume. To accomplish this the scattered light from each red cell in the sample is measured in real time at two angular intervals. The cell volume and the hemoglobin concentration within the cell are derived from these two measurements. Severe accuracy and precision specifications are placed on the medically important red cell count (RBC) and the mean red cell volume (MCV). From the point of view of optical system design, the dominant factor is the requirement that RBC and MCV have precision and accuracy of the order of 2%. Signal-to-noise and scattering-angle definition requirements dictated the choice of a HeNe laser light source. The optics includes an illumination system for producing a sharply defined, uniformly illuminated scattering region and a detection system which must accurately define the accepted scattering angles. In previous cytometric methods for determining MCV only a single quantity was measured for each cell. Such methods cannot disentangle the independent effects of cell size and hemoglobin concentration on the measurement, thus compromising MCV accuracy. The present double-angle scattering method overcomes this accuracy problem. The H*1 red cell method, the supporting optical design and data demonstrating that the use of this technique eliminates interference between the observed red cell indices are presented.
© (1988) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
G M Colella, D H Tycko, and W Groner "Technicon H*1 Hematology System: Optical Design Considerations", Proc. SPIE 0907, Laser Surgery: Characterization and Therapeutics, (15 June 1988); https://doi.org/10.1117/12.945330
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Light scattering

Blood

Laser therapeutics

Refractive index

Hematology

Optical design

Scattering

RELATED CONTENT

Physical principles of monolithic high-contrast gratings
Proceedings of SPIE (February 20 2017)
Small-angle scattering measurement
Proceedings of SPIE (November 11 1991)
Critical point facility for the IML 2 mission reduction...
Proceedings of SPIE (September 30 1994)
Panel Discussion - Measurements & Standards
Proceedings of SPIE (September 26 1977)

Back to Top