20 December 2016 Gain drift compensation with no feedback-loop developed for the X-Ray Integral Field Unit/ATHENA readout chain
Damien Prêle, Fabrice Voisin, Cyril Beillimaz, Si Chen, Andrea Goldwurm
Author Affiliations +
Abstract
The focal plane of the X-Ray Integral Field Unit (X-IFU) instrument of the Advanced Telescope for High-Energy Astrophysics observatory is composed of 3840 microcalorimeters. These sensors, based on superconducting transition edge sensors (TES), are read out through a frequency multiplexer. A “base-band feedback” suppresses all the carriers of the multiplexed signal in the superconducting quantum interference devices input coil (cryogenic readout). However, the loop gain of this feedback is too small (less than 10 in the present baseline of the phase A mission) to strongly compensate the readout gain drifts. An onboard x-ray source is considered to calibrate the gain of the full instrument. However, in-flight calibration time must be minimized, which leads to a requirement on the gain stability larger than 104 over a long duration (between each calibration) to reach the challenging energy resolution goal of 2.5 eV at 6 keV of the X-IFU. A significant part of this gain is provided by a low-noise amplifier in the warm front-end electronics (WFEE). To reach such gain stability over more than a dozen minutes, this noncooled amplifier has to cope with the temperature and supply voltage variations. Moreover, mainly for noise reasons, a common large loop gain with feedback cannot be used. We propose a new amplifier topology using diodes as loads of a differential amplifier to provide a fixed voltage gain, independent of the temperature and of the bias fluctuations. This amplifier is designed using 350-nm SiGe BiCMOS technology and is part of an integrated circuit developed for the WFEE. Our simulations provide the expected gain and noise performances. Comparison with standard resistive loaded differential pair clearly shows the advantages of the proposed amplifier topology with a gain drift decreased by more than an order of magnitude. Performances of this diode loaded amplifier are discussed in the context of the X-IFU requirements.
© 2016 Society of Photo-Optical Instrumentation Engineers (SPIE) 2329-4124/2016/$25.00 © 2016 SPIE
Damien Prêle, Fabrice Voisin, Cyril Beillimaz, Si Chen, and Andrea Goldwurm "Gain drift compensation with no feedback-loop developed for the X-Ray Integral Field Unit/ATHENA readout chain," Journal of Astronomical Telescopes, Instruments, and Systems 2(4), 046002 (20 December 2016). https://doi.org/10.1117/1.JATIS.2.4.046002
Received: 11 July 2016; Accepted: 18 November 2016; Published: 20 December 2016
Lens.org Logo
CITATIONS
Cited by 6 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Amplifiers

Diodes

Transistors

X-ray telescopes

X-rays

Resistance

Multiplexing

RELATED CONTENT

The gas pixel detector at the focus of an x...
Proceedings of SPIE (September 26 2013)
X-IFU technical challenge
Proceedings of SPIE (July 15 2016)
Absolute timing with the SWIFT X-ray telescope (XRT)
Proceedings of SPIE (August 18 2005)
VUV And Soft X Ray Radiometry At The Laboratory Of...
Proceedings of SPIE (December 21 1988)
Silicon photodiodes for absolute soft x-ray radiometry
Proceedings of SPIE (September 20 2007)

Back to Top