1.IntroductionNumerous lung diseases, such as asthma, may impact the mechanical behavior of the airway walls.1, 2, 3 Airway remodeling, commonly observed in asthmatic patients, involves thickening of the mucosa, collagen deposition, and disruption of the normal elastin network, which can affect the tissue mechanics.4, 5, 6, 7, 8, 9 As airway smooth muscle contracts in parallel with the noncontractile tissue, changes in the mechanical properties of the tissue could affect smooth muscle contractility and thus the pathogenesis of asthma.3 Assessing airway tissue mechanics in vivo could improve the understanding, diagnosis, and management of lung diseases such as asthma, which affect at least 5% of the U.S. adult population.10 The in vivo mechanical behavior of the airways remains poorly understood, as current techniques lack the temporal-spatial resolution necessary to track their deformation during normal tidal breathing. Current techniques to evaluate tissue compliance are either ex vivo techniques, which require excision and thus tissue disruption,11, 12 or in vivo techniques which either rely on contact with the tissue, loading the airways with nonphysiologic methods,2 or rely on imaging methods which lack the temporal resolution to measure deformations occurring within a single breath.13 It is known that the smooth muscle surrounding the airways (the primary effector of airway narrowing in asthma) is acutely sensitive to strain and strain rate, suggesting that the ideal assessment of in vivo airway mechanics would be direct, noncontact measure of airway strain during normal, tidal breathing. In order to track the small deformations occurring in the large airways during normal tidal breathing, an imaging modality with high spatial resolution (submillimeter), and high temporal resolution (>10 Hz) is needed. Optical coherence tomography (OCT) is an emerging medical imaging technique which allows for high resolution (∼10 μm/pixel) images of tissue. Endoscopic probes have been developed for many organs, allowing imaging of the blood vessels, esophagus, urethra, and airways with a resolution previously only available via biopsy and histological preparation. 14, 15, 16, 17, 18, 19, 20, 21 Recently developed Fourier-domain OCT techniques maintain the same spatial resolution while offering video-frame rates (up to 100 frames/s).20, 21 Furthermore, unlike ultrasound, OCT can be used in noncontact modes and in air-filled lumens, such as the airways. Thus, the speed and resolution of OCT make it potentially ideal for mechanical property assessment of the airways. 22, 23, 24, 25, 26, 27, 28, 29, 30, 31 In order for OCT-based mechanical analysis of airways to be possible, algorithms adaptable to endoscopic probes are needed, as are ways to overcome confounding motion artifacts. One previous work took OCT images of the airways at end expiration with different levels of positive end expiratory pressure (PEEP), however this technique neither evaluated motion at respiration frequency nor was able to independently analyze different regions of the trachea.13 While elastography methods which use cross correlation techniques have been published, these techniques are difficult to adapt to flexible probes.31 Similarly, Doppler-based techniques are sensitive to probe motion as the tissue to be imaged and the endoscopic probe both move; thus, techniques which are robust to probe motion are needed.32, 33 We have developed a new algorithm for mechanical analysis of tissue using OCT in vivo, deemed OCT wall motion analysis (OCT-WMA). This approach offers a simple, robust estimate of the mechanical response of the airways to tidal breathing in vivo. 2.Methods2.1.OCT System and Image AcquisitionThe OCT system used for this study has been previously described.19, 21 OCT image stacks were acquired using a custom-built Fourier-domain system with an endoscopic probe, which was constructed with a rotating microelectromechanical motor (Namiki Prevision Jewel Co. Ltd., Tokyo, Japan), a single mode optical fiber gradient index lens (NSG America Inc., Somerset, New Jersey), and a 45º prism mirror (Tower Optical Corporation, Boynton Beach, Florida). This system uses a swept laser (Santec Corporation, Aichi, Japan) at 1310 nm with a full-width-half-maximum of 100 nm, giving a theoretical axial resolution of 7.5 μm. The axial scan range was 3.5 mm in air and the sensitivity of this system was measured to be 107 dB with a 6-dB roll-off at the depth of 2.2 mm. The measured axial and lateral resolutions in air were 8 and 20 μm, respectively. During tidal breathing, 400 two-dimensional (2D) images were taken of the same axial location with 1024 (lateral) × 400 (axial) pixel size at a rate of 19.5 frames/s (i.e., approximately 20 s of images of the same axial location) generating an image set for mechanical analysis (Video 1). Ventilatory pressure at the inlet and at the imaging site were measured concurrently with a 1 PSI transducer (Honeywell, Golden Valley, Minnesota) and a 1F catheter pressure transducer (Millar Instruments, Houston, Texas), respectively. 10.1117/1.3642006.12.2.Algorithm DevelopmentIn order to calibrate distance in pixels from the OCT images and distance in millimeters, images were made in air of the inside of tubes of known radius, generating a linear (r 2 > 0.99) calibration curve with a slope of 0.0103 mm/pixel in the radial direction. The surface of the tissue was identified in a semiautomated fashion via custom image processing scripts written in MATLAB (Mathworks, Natick, Massachusetts). As the swept-source OCT images used in this study contained a large number of artifacts due to internal reflections, specifying a search region in the image was necessary for accurate recognition of the tissue surface. Our algorithm has the user select 20 points along the surface in the first image of a 400-image set; then a curve is interpolated through these points via cubic splines. This spline curve serves to define the center of the search region, and the tissue surface in each vertical line is detected via a combination of gradient and intensity. The tissue surface in the first image slice then serves to define the search region in the next image of the set, and the process repeats through the entire 400-image set. By taking the surface of the tissue in each frame, the location of the tissue surface in space and time can be tracked. The repeatability of this algorithm was established via repeated (selecting a different set of initial 20 random points) analysis of the same image set. The average error between runs (with two different users) was 0.005 pixels/vertical scan line for six runs. In order to reduce the effect of probe motion, a circle was fit through the tissue surface in each image. The center of these circles was then co-localized and the surfaces adjusted accordingly. As cardiac motion and other sources of noise generate unwanted deformations in the tissue, the next step in the analysis was to remove motion at frequencies other than the driving frequency of tidal breathing. At each radial location, respiratory motion was identified via a Fourier technique: the distance from probe center to tissue surface in time was extracted for each position around the trachea; then the Fourier transform of this data series was taken. The magnitude and phase of the peak at respiration frequency were extracted. This value is proportional to the peak-to-peak height of the motion due to respiration: If the series x k is a discretized sinusoid, with amplitude A and period h/N, then the Fourier transform at frequency h is
[TeX:]
\documentclass[12pt]{minimal}\begin{document}\begin{equation*}
F(h) = F(- h) = \sum\limits_{n = 0}^{N - 1} {\frac{A}{2}\,} e^{(2\pi i/N)\,hn} e^{(- 2\pi i/N)\,hn} = \frac{{A\,N}}{2}.
\end{equation*}\end{document}
Therefore, to extract the amplitude at this frequency, we multiplied the discrete Fourier transform at the desired frequency by 2/N. This value can be calculated for each angular position in the image, representing the peak-to-peak radial deformation in the tissue due to respiration at a given position in the trachea. The local radial deformation, measured by this technique, was then synced in time to a pressure tracing in order to give apparent local wall compliance. Radial strain (ɛ) was defined as the change in the inner radius (Δr) normalized to radius (r) which is related to the change in the arc length (ΔC) of the trachea divided by the original arc length (C):
Eq. 1[TeX:] \documentclass[12pt]{minimal}\begin{document}\begin{equation} \varepsilon = \frac{{\Delta r}}{r} = \frac{{\Delta C}}{C}. \end{equation}\end{document}Eq. 2[TeX:] \documentclass[12pt]{minimal}\begin{document}\begin{equation} \sigma \times d = P \times r. \end{equation}\end{document}Using the thin walled tube equation, Eqs. 1, 2 can be rearranged to provide an expression for wall compliance, W c. Eq. 3[TeX:] \documentclass[12pt]{minimal}\begin{document}\begin{equation} W_c = \frac{\varepsilon }{{\sigma \times d}} = \frac{{\Delta r/r}}{{P \times r}} = \frac{1}{{E \times d}}. \end{equation}\end{document}3.Results3.1.Validation Using PhantomsOur algorithm was validated by imaging and analyzing the material properties of elastic tube phantoms with both the proposed OCT WMA scheme and standard tensile mechanical testing. Two different phantoms were created, a uniform elastic tube and a tube with different compliances in the circumferential direction. First, the uniform tube-shaped elastic phantom was inflated using a small animal ventilator (Harvard Apparatus, Holliston, Massachusetts) during imaging, while the ventilatory pressure was recorded using a 12bit A/D system (I-158, DATAQ, Akron, Ohio) with a calibrated 1 psi transducer at the inlet (Honeywell, Morristown, New Jersey) and a 1F catheter pressure transducer at the imaging site (Millar Instruments, Houston, Texas). The same material was then cut into 6-mm wide × 2-cm long strips and tested in tensile loading from 0 to 78 kPa of stress (equivalent to the stress range during tidal inflation) at a strain rate of 1 mm/min on a uniaxial test frame (MTS Synergie 100, MTS, Eden Prairie, Minnesota). Materials testing of the phantom (tangential modulus- 1.3±0.3 MPa) gives a W c of 6.0±1.3 × 10−6 (mm Pa)−1. The wall compliance calculated from OCT imaging was 4.5±2 ×10−6 mm Pa−1 if using the pressure sensor at the inlet and 6.3±0.3 ×10−6 mm Pa−1 if using the sensor at the imaging site (Fig. 1). Next, we tested our algorithm in a phantom with a rigid (thick PVC tubing) and a compliant section (thin rubber) to model the trachea (rigid C-shaped cartilage rings attached by the compliant trachealis muscle). We were readily able to discern the compliant section via these methods, even though we could not directly image the rigid tubing from these images (Fig. 2). 3.2.In Vivo Wall Motion Analysis of Rabbit TracheaMechanical analysis using OCT was performed in vivo in a rabbit model. The animals (male adult New Zealand White, 4 to 4.5 kg, Western Oregon Rabbit Co. Philomath, Oregon) were anesthetized with a ketamine-xylazine mixture, intubated with a 3.5 mm endotracheal tube, and imaged with our OCT system as described above and previously.11, 15, 16 All methods were approved by the Institutional Animal Care and Use Committee at the University of California, Irvine (Protocol No. 2007-2754). The OCT probe was inserted into the trachea of the animal through the intubation tube, and the tip positioned approximately 1 cm beyond the distal end of the intubation tube within the lower trachea. An image set was acquired (as described in Sec. 2.2) while the animal was ventilated during normal tidal breathing (Video 1). The image set was processed as described above in order to suppress motion artifacts (e.g., probe motion, cardiac motion) and calculate local wall compliance (Fig. 3) By stacking images at different axial locations along the trachea, a compliance map of the trachea can be generated (Fig. 4). The mean tracheal stiffness across polar coordinate and longitudinal distance was approximately 1.3±0.3×10−5 mm Pa−1. This value gives a tangential modulus of 25 to 75 kPa (assuming a range of tissue thickness of 1 to 3 mm) which differs appreciably from the previously reported ex vivo modulus of 4.1 kPa.11 Increased compliance is notable in the region corresponding to the trachealis, and closer to the carina. 4.DiscussionEvaluating the mechanical properties of the airways during normal ventilatory function may aid in the understanding of pulmonary diseases ranging from asthma to sleep apnea; however, previous methods have not been able to capture in vivo airway mechanical behavior with normal breathing generating the displacements. Using OCT to image the airway wall, we have developed algorithms to extract airway wall motion during normal, tidal ventilation, calculate wall compliance, validated these techniques against phantoms, and used them to calculate wall compliance in vivo. Since OCT has higher spatial and temporal resolution than previously used techniques such as ultrasound or magnetic resonance, this technique presents a potential advance for in vivo analysis of tissue mechanics. This technique is distinct from previously described elastography algorithms, as it tracks the surface of the tissue in time, as opposed to using cross correlation to track movement within the tissue.23, 24, 34 This offers improved signal-to-noise, but loses the ability to track different tissue layers. Unlike previous works which investigated tissue mechanics by measuring relative strain,35, 36 we directly measured both strain and driving pressure, allowing for quantitative analysis of the airway wall mechanics and comparison to other mechanical testing methods. Additionally, we have demonstrated techniques which function at breathing frequency and localize motion to different regions of the trachea: significant advances compared to previous in vivo work.13 Validation with phantoms showed good accordance between the methods developed in this work and standard mechanical testing. A tissue thickness range of 1 to 3 mm [Eq. 3], our value for W c gives a tangential modulus of approximately 25 to 75 kPa, which is larger than the previously reported ex vivo modulus of 4.1 kPa.11 This difference may be due to active smooth muscle tension during normal loading in vivo.37 This technique does have limitations. The probe size is large relative to the lumen diameter of the endotracheal tube, which could distort the air flow pattern; however, we have attempted to mitigate these effects by measuring the pressure at the imaging site. Similarly, given the size of our OCT probe we could not image smaller airways in a rabbit model, though the size of the probe would allow imaging of human airways down to generation #5 (diameter > 3 mm).38 Additionally, the technique developed in this work calculates compliance based on radial deformation, ignoring deformations in the circumferential direction. It is possible that in inhomogeneous materials we miss some purely circumferential deformation even though increases in circumference are linked to increases in radius [by Eq. 1]. Finally, the deformations of the wall surface represent a composite for all the tissue under the surface; thus, our technique cannot assess the mechanical properties of specific tissue layers (e.g., smooth muscle alone). Furthermore, for the analysis of relative compliance, the stress in the wall was considered to be homogeneous (thin wall assumption), which is not an accurate representation of the unknown (but undoubtedly complex) stress distribution in the wall. This assumption was used as the thickness of the airway wall cannot be directly measured from the OCT images (as the distance in pixels depends on the index of refraction of the mucosa- currently unknown) and to simplify the mathematics to an easily understood and simply computed model. These assumptions undoubtedly oversimplify the problem; however, they reduce an essentially unsolvable system to interesting metrics of strain in various regions of the tissue. Unlike previous techniques, our algorithm considers dynamic changes in driving pressure instead of measuring only relative image strain. Even though our method is simpler than many previously suggested algorithms, it remains robust to noise. Several previous reports have used cross correlation-based techniques to investigate tissue mechanics; 22, 23, 25, 26, 27, 39, 40 however, we found that with probe and cardiac motion accounting for as much as 50% of the observed deformation, alternative techniques were needed. Similarly, Doppler-based techniques work best when unwanted motion is considerably slower and of smaller magnitude than the deformation of interest. For example, blood flow is easily quantified this way.41, 42 In contrast, the cardiac motion artifact from this study was faster and of similar magnitude to the respiratory motion, yet our technique of Fourier transform and bandpass filtering was able to reject this unwanted motion artifact without compromising accuracy. 5.ConclusionsWe have developed a new OCT-based method to analyze the mechanical properties of airways. By comparing the deflection of the surface of the tissue to ventilatory pressure during tidal breathing we have evaluated in vivo mechanical behavior (i.e., wall compliance) of a rabbit trachea. The high temporal and spatial resolution of OCT makes this technique more sensitive than previously reported techniques, and concurrent pressure monitoring allows for the estimation of the absolute compliance. This technique may have applications in research and clinical diagnostics and may facilitate our understanding of in vivo airway dynamics of prevalent lung diseases such as asthma that are characterized by change in tissue mechanics. AcknowledgmentsWe would like to thank David Mukai, Tanya Burney, David Yoon, Kenji Ikemura, Emerald Chun, and Anna Aledia for their contributions. We would also like to acknowledge our funding sources including NIH 5R01CA124967 and NIH R01HL067954. Claire Robertson is supported by the NSF Graduate Research Fellowship Program. ReferencesN. J. Brown,
C. M. Salome,
N. Berend,
C. W. Thorpe, and
G. G. King,
“Airway distensibility in adults with asthma and healthy adults, measured by forced oscillation technique,”
Am. J. Respir. Crit. Care Med., 176
(2), 129
–137
(2007). https://doi.org/10.1164/rccm.200609-1317OC Google Scholar
H. J. Brackel,
O. F. Pedersen,
P. G. Mulder,
S. E. Overbeek,
K. F. Kerrebijn, and
J. M. Bogaard,
“Central airways behave more stiffly during forced expiration in patients with asthma,”
Am. J. Respir. Crit. Care Med., 162
(3 Pt 1), 896
–904
(2000). Google Scholar
A. M. Bramley,
R. J. Thomson,
C. R. Roberts, and
R. R. Schellenberg,
“Hypothesis: excessive bronchoconstriction in asthma is due to decreased airway elastance,”
Eur. Respir. J., 7
(2), 337
–341
(1994). https://doi.org/10.1183/09031936.94.07020337 Google Scholar
Y. Bosse,
P. D. Pare, and
C. Y. Seow,
“Airway wall remodeling in asthma: from the epithelial layer to the adventitia,”
Curr. Allergy Asthma Rep., 8
(4), 357
–366
(2008). https://doi.org/10.1007/s11882-008-0056-0 Google Scholar
W. Busse,
J. Elias,
D. Sheppard, and
S. Banks-Schlegel,
“Airway remodeling and repair,”
Am. J. Respir. Crit. Care Med., 160
(3), 1035
–1042
(1999). Google Scholar
F. H. Chen,
K. T. Samson,
K. Miura,
K. Ueno,
Y. Odajima,
T. Shougo,
Y. Yoshitsugu, and
S. Shioda,
“Airway remodeling: a comparison between fatal and nonfatal asthma,”
J. Asthma, 41
(6), 631
–638
(2004). https://doi.org/10.1081/JAS-200026405 Google Scholar
J. A. Elias,
“Airway remodeling in asthma. Unanswered questions,”
Am. J. Respir. Crit. Care Med., 161
(3 Pt 2), S168
–S171
(2000). Google Scholar
J. A. Elias,
Z. Zhu,
G. Chupp, and
R. J. Homer,
“Airway remodeling in asthma,”
J. Clin. Invest., 104
(8), 1001
–1006
(1999). https://doi.org/10.1172/JCI8124 Google Scholar
B. Fabry and
J. J. Fredberg,
“Remodeling of the airway smooth muscle cell: are we built of glass,”
Respir. Physiol. Neurobiol., 137
(2–3), 109
–124
(2003). https://doi.org/10.1016/S1569-9048(03)00141-1 Google Scholar
T. M. Eagan,
J. C. Brogger,
G. E. Eide, and
P. S. Bakke,
“The incidence of adult asthma: a review,”
Int. J. Tuberc. Lung Dis., 9
(6), 603
–612
(2005). Google Scholar
C. B. Raub,
S. Mahon,
N. Narula,
B. J. Tromberg,
M. Brenner, and
S. C. George,
“Linking optics and mechanics in an in vivo model of airway fibrosis and epithelial injury,”
J. Biomed. Opt., 15
(1), 015004 https://doi.org/10.1117/1.3322296 Google Scholar
P. R. Cooper,
B. E. McParland,
H. W. Mitchell,
P. B. Noble,
A. Z. Politi,
A. R. Ressmeyer, and
A. R. West,
“Airway mechanics and methods used to visualize smooth muscle dynamics in vitro,”
Pulm. Pharmacol. Ther., 22
(5), 398
–406
(2009). https://doi.org/10.1016/j.pupt.2008.09.008 Google Scholar
J. P. Williamson,
R. A. McLaughlin,
W. J. Noffsinger,
A. L. James,
V. A. Baker,
A. Curatolo,
J. J. Armstrong,
A. Regli,
K. L. Shepherd,
G. B. Marks,
D. D. Sampson,
D. R. Hillman, and
P. R. Eastwood,
“Elastic properties of the central airways in obstructive lung diseases measured using anatomical optical coherence tomography,”
Am. J. Respir. Crit. Care Med., 183 612
–619
(2011). Google Scholar
J. Armstrong,
M. Leigh,
I. Walton,
A. Zvyagin,
S. Alexandrov,
S. Schwer,
D. Sampson,
D. Hillman, and
P. Eastwood,
“In vivo size and shape measurement of the human upper airway using endoscopic longrange optical coherence tomography,”
Opt. Express, 11
(15), 1817
–1826
(2003). https://doi.org/10.1364/OE.11.001817 Google Scholar
M. Brenner,
K. Kreuter,
J. Ju,
S. Mahon,
L. Tseng,
D. Mukai,
T. Burney,
S. Guo,
J. Su,
A. Tran,
A. Batchinsky,
L. C. Cancio,
N. Narula, and
Z. Chen,
“In vivo optical coherence tomography detection of differences in regional large airway smoke inhalation induced injury in a rabbit model,”
J. Biomed. Opt., 13
(3), 034001
(2008). https://doi.org/10.1117/1.2939400 Google Scholar
M. Brenner,
K. Kreuter,
D. Mukai,
T. Burney,
S. Guo,
J. Su,
S. Mahon,
A. Tran,
L. Tseng,
J. Ju, and
Z. Chen,
“Detection of acute smoke-induced airway injury in a New Zealand white rabbit model using optical coherence tomography,”
J. Biomed. Opt., 12
(5), 051701
(2007). https://doi.org/10.1117/1.2798637 Google Scholar
S. Han,
N. H. El-Abbadi,
N. Hanna,
U. Mahmood,
R. Mina-Araghi,
W. G. Jung,
Z. Chen,
H. Colt, and
M. Brenner,
“Evaluation of tracheal imaging by optical coherence tomography,”
Respiration, 72
(5), 537
–541
(2005). https://doi.org/10.1159/000087680 Google Scholar
N. Hanna,
D. Saltzman,
D. Mukai,
Z. Chen,
S. Sasse,
J. Milliken,
S. Guo,
W. Jung,
H. Colt, and
M. Brenner,
“Two-dimensional and 3-dimensional optical coherence tomographic imaging of the airway, lung, and pleura,”
J. Thorac. Cardiovasc. Surg., 129
(3), 615
–622
(2005). https://doi.org/10.1016/j.jtcvs.2004.10.022 Google Scholar
J. Su,
J. Zhang,
L. Yu,
G. C. H,
M. Brenner, and
Z. Chen,
“Real-time swept source optical coherence tomography imaging of the human airway using a microelectromechanical system endoscope and digital signal processor,”
J. Biomed. Opt., 13
(3), 030506
(2008). https://doi.org/10.1117/1.2938700 Google Scholar
T. Xie,
G. Liu,
K. Kreuter,
S. Mahon,
H. Colt,
D. Mukai,
G. M. Peavy,
Z. Chen, and
M. Brenner,
“In vivo three-dimensional imaging of normal tissue and tumors in the rabbit pleural cavity using endoscopic swept source optical coherence tomography with thoracoscopic guidance,”
J. Biomed. Opt., 14
(6), 064045
(2009). https://doi.org/10.1117/1.3275478 Google Scholar
J. Yin,
G. Liu,
J. Zhang,
L. Yu,
S. Mahon,
D. Mukai,
M. Brenner, and
Z. Chen,
“In vivo early detection of smoke-induced airway injury using three-dimensional swept-source optical coherence tomography,”
J. Biomed. Opt., 14
(6), 060503
(2009). https://doi.org/10.1117/1.3268775 Google Scholar
S. G. Adie,
B. F. Kennedy,
J. J. Armstrong,
S. A. Alexandrov, and
D. D. Sampson,
“Audio frequency in vivo optical coherence elastography,”
Phys. Med. Biol., 54
(10), 3129
–3139
(2009). https://doi.org/10.1088/0031-9155/54/10/011 Google Scholar
R. Chan,
A. Chau,
W. Karl,
S. Nadkarni,
A. Khalil,
N. Iftimia,
M. Shishkov,
G. Tearney,
M. Kaazempur-Mofrad, and
B. Bouma,
“OCT-based arterial elastography: robust estimation exploiting tissue biomechanics,”
Opt. Express, 12
(19), 4558
–4572
(2004). https://doi.org/10.1364/OPEX.12.004558 Google Scholar
B. F. Kennedy,
T. R. Hillman,
R. A. McLaughlin,
B. C. Quirk, and
D. D. Sampson,
“In vivo dynamic optical coherence elastography using a ring actuator,”
Opt. Express, 17
(24), 21762
–21772
(2009). https://doi.org/10.1364/OE.17.021762 Google Scholar
A. S. Khalil,
R. C. Chan,
A. H. Chau,
B. E. Bouma, and
M. R. Mofrad,
“Tissue elasticity estimation with optical coherence elastography: toward mechanical characterization of in vivo soft tissue,”
Ann. Biomed. Eng., 33
(11), 1631
–1639
(2005). https://doi.org/10.1007/s10439-005-6766-3 Google Scholar
S. J. Kirkpatrick,
R. K. Wang, and
D. D. Duncan,
“OCT-based elastography for large and small deformations,”
Opt. Express, 14
(24), 11585
–11597
(2006). https://doi.org/10.1364/OE.14.011585 Google Scholar
H. J. Ko,
W. Tan,
R. Stack, and
S. A. Boppart,
“Optical coherence elastography of engineered and developing tissue,”
Tissue Eng., 12
(1), 63
–73
(2006). https://doi.org/10.1089/ten.2006.12.63 Google Scholar
X. Liang and
S. Boppart,
“Biomechanical properties of in vivo human skin from dynamic optical coherence elastography,”
IEEE Trans. Biomed. Eng., 57
(4), 953
–959
(2009). Google Scholar
J. Rogowska,
N. A. Patel,
J. G. Fujimoto, and
M. E. Brezinski,
“Optical coherence tomographic elastography technique for measuring deformation and strain of atherosclerotic tissues,”
Heart, 90
(5), 556
–562
(2004). https://doi.org/10.1136/hrt.2003.016956 Google Scholar
J. Schmitt,
“OCT elastography: imaging microscopic deformation and strain of tissue,”
Opt. Express, 3
(6), 199
–211
(1998). https://doi.org/10.1364/OE.3.000199 Google Scholar
G. van Soest,
F. Mastik,
N. de Jong, and
A. F. van der Steen,
“Robust intravascular optical coherence elastography by line correlations,”
Phys. Med. Biol., 52
(9), 2445
–2458
(2007). https://doi.org/10.1088/0031-9155/52/9/008 Google Scholar
J. Walther,
A. Kruger,
M. Cuevas, and
E. Koch,
“Effects of axial, transverse, and oblique sample motion in FD OCT in systems with global or rolling shutter line detector,”
J. Opt. Soc. Am. A, 25
(11), 2791
–2802
(2008). https://doi.org/10.1364/JOSAA.25.002791 Google Scholar
Z. Xu,
L. Carrion, and
R. Maciejko,
“A zero-crossing detection method applied to Doppler OCT,”
Opt. Express, 16
(7), 4394
–4412
(2008). https://doi.org/10.1364/OE.16.004394 Google Scholar
R. Karimi,
T. Zhu,
B. E. Bouma, and
M. R. Mofrad,
“Estimation of nonlinear mechanical properties of vascular tissues via elastography,”
Cardiovasc. Eng., 8
(4), 191
–202
(2008). https://doi.org/10.1007/s10558-008-9061-0 Google Scholar
E. I. Cespedes,
C. L. de Korte, and
A. F. van der Steen,
“Intraluminal ultrasonic palpation: assessment of local and cross-sectional tissue stiffness,”
Ultrasound Med. Biol., 26
(3), 385
–396
(2000). https://doi.org/10.1016/S0301-5629(99)00169-6 Google Scholar
M. M. Doyley,
F. Mastik,
C. L. de Korte,
S. G. Carlier,
E. I. Cespedes,
P. W. Serruys,
N. Bom, and
A. F. van der Steen,
“Advancing intravascular ultrasonic palpation toward clinical applications,”
Ultrasound Med. Biol., 27
(11), 1471
–1480
(2001). https://doi.org/10.1016/S0301-5629(01)00457-4 Google Scholar
S. S. An,
T. R. Bai,
J. H. Bates,
J. L. Black,
R. H. Brown,
“Airway smooth muscle dynamics: a common pathway of airway obstruction in asthma,”
Eur. Respir. J., 29
(5), 834
–860
(2007). https://doi.org/10.1183/09031936.00112606 Google Scholar
E. R. Weibel,
“Morphometry of the human lung: the state of the art after two decades,”
Bull. Eur. Physiopathol. Respir., 15
(5), 999
–1013
(1979). Google Scholar
R. A. Baldewsing,
J. A. Schaar,
F. Mastik, and
A. F. van der Steen,
“Local elasticity imaging of vulnerable atherosclerotic coronary plaques,”
Adv. Cardiol., 44 35
–61
(2007). https://doi.org/10.1159/000096719 Google Scholar
Y. Li and
J. G. Snedeker,
“Elastography: modality-specific approaches, clinical applications, and research horizons,”
Skeletal. Radiol., 40
(4), 389
–397
(2011). Google Scholar
A. Davis,
J. Izatt, and
F. Rothenberg,
“Quantitative measurement of blood flow dynamics in embryonic vasculature using spectral Doppler velocimetry,”
Anat. Rec., 292
(3), 311
–319
(2009). https://doi.org/10.1002/ar.20808 Google Scholar
A. Mariampillai,
B. A. Standish,
E. H. Moriyama,
M. Khurana,
N. R. Munce,
M. K. Leung,
J. Jiang,
A. Cable,
B. C. Wilson,
I. A. Vitkin, and
V. X. Yang,
“Speckle variance detection of microvasculature using swept-source optical coherence tomography,”
Opt. Lett., 33
(13), 1530
–1532
(2008). https://doi.org/10.1364/OL.33.001530 Google Scholar
|