2 March 2017 Smartphone snapshot mapping of skin chromophores under triple-wavelength laser illumination
Janis Spigulis, Ilze Oshina, Anna Berzina, Alexander Bykov
Author Affiliations +
Abstract
Chromophore distribution maps are useful tools for skin malformation severity assessment and for monitoring of skin recovery after burns, surgeries, and other interactions. The chromophore maps can be obtained by processing several spectral images of skin, e.g., captured by hyperspectral or multispectral cameras during seconds or even minutes. To avoid motion artifacts and simplify the procedure, a single-snapshot technique for mapping melanin, oxyhemoglobin, and deoxyhemoglobin of in-vivo skin by a smartphone under simultaneous three-wavelength (448–532–659 nm) laser illumination is proposed and examined. Three monochromatic spectral images related to the illumination wavelengths were extracted from the smartphone camera RGB image data set with respect to crosstalk between the RGB detection bands. Spectral images were further processed accordingly to Beer’s law in a three chromophore approximation. Photon absorption path lengths in skin at the exploited wavelengths were estimated by means of Monte Carlo simulations. The technique was validated clinically on three kinds of skin lesions: nevi, hemangiomas, and seborrheic keratosis. Design of the developed add-on laser illumination system, image-processing details, and the results of clinical measurements are presented and discussed.

1.

Introduction

Abnormal distribution of skin absorbing chromophores (e.g., epidermal melanin and dermal hemoglobin) is a clinical feature typical for burns, bruises, and various kinds of skin malformations, including tumors.1 Skin chromophore distribution maps can provide qualitative and quantitative information on the severity of pathology, its development, and/or skin recovery processes. Simple and fast skin chromophore mapping is of special interest for dermatologists, cosmetologists, oncologists, emergency clinicians, surgeons, and forensic experts, as it facilitates decision taking and further strategy development. However, the procedures with commercially available devices that can provide skin chromophore distribution maps (e.g., SIAscope)2 are time-consuming, the equipment is expensive and is cable-connected to a computer, which limits its wide applications in personalized medicine and in the field conditions.

Each chromophore has its specific absorption spectrum,3 so the remitted light intensity from a spot of skin (containing one or more chromophores) depends on the detection wavelength and composition of the comprised chromophores. A set of several spectral images of the same skin area, captured within different wavelength bands, can be converted into clinically significant parametric images—chromophore distribution maps.4 There are two main approaches for obtaining spectral images: by narrowband spectral filtering of the image sensor at spectrally broadband illumination of the target area or by means of spectrally narrowband illumination. The first approach is widely used in hyperspectral imaging (HSI) systems that capture sequential images of the target at several tens of adjacent spectral bands, e.g., using interference filters,5 acousto-optical, or liquid crystal filters.6 Commercial and custom-designed HSI systems have been successfully used for skin chromophore mapping.5,79 However, routine clinical applications of the HSI systems are limited due to their robustness, expensiveness, and the need for powerful computing resources to process numerous spectral images (including those eventually not containing useful information). Another advanced chromophore-sensitive technology, spatial frequency domain spectroscopy,9 faces similar limitations. The HSI image acquisition process may last from several seconds9 to a minute or longer,10 so movements of the skin target area during the procedure can cause image artifacts. Special image stabilization algorithms11 could reduce or fully avoid the motion artifacts, but it takes extra time and computing resources. Additional problems arise if the extinction coefficient of a particular chromophore is notably changing within a single band wavelength interval, typically 10 to 20 nm or more; in such cases, calculations of complicated integrals comprising spectral features of the chromophore absorbance and the filtered HSI band shape are necessary.

The second approach, spectrally selective illumination, has been mainly used in skin multispectral imaging (MSI) systems where sequentially switchable nonoverlapping spectral bands of illumination (e.g., by several LEDs with different peak wavelengths at typical spectral half-bandwidth 20 to 50 nm) are exploited for obtaining spectral images.12,13 Advanced MSI systems are portable and self-sustained, much cheaper than HSI systems, and better adapted for clinical use. The MSI processing procedure is also faster due to lower number of the spectral images, typically between 3 and 8. Still, skin motion artifacts and processing of relatively broadband spectral images may cause serious problems.

To improve the performance of the MSI chromophore mapping, acquisition time of the spectral images and the spectral bandwidth of illumination have to be minimized. An optimal approach would be the single-snapshot technique dealing with monochromatic (single-wavelength) spectral images. A method for extraction of three monochromatic spectral images from a single snapshot RGB image data set, taken under simultaneous three-wavelength illumination, has been recently proposed14 and experimentally validated on color targets.15 The previous studies confirmed that separate readings of three RGB detection channels are sufficient to spectrally isolate reflected light at each illumination wavelength, if the crosstalk among those channels is properly taken into account.16 A table-top setup comprising RGB board-camera and fiber optic system for simultaneous illumination of target by the 473-, 532-, and 659-nm laser lines has been tested for skin chromophore mapping, as well.17 In the results, physiologically plausible chromophore maps of melanin, oxyhemoglobin, and deoxyhemoglobin in pigmented and vascular skin lesions have been obtained by a single snapshot. This laboratory setup, however, was not well-adapted for patient measurements, so new designs of compact/portable equipment for skin chromophore mapping in clinical environment are of essential interest.

Currently, there are about two billion smartphones in use worldwide,18 many of them in the pockets of doctors or other clinical personnel. Smartphones comprise high-resolution digital RGB cameras, white LED illuminators, multifunctional displays, and powerful data processors with the ability to transmit the data (including specific images) via mobile networks. This set of features makes smartphones attractive for potential diagnostic applications, e.g., evaluation of human skin condition. Several dermatology-specific mobile applications, including those intended for skin melanoma early detection, have been developed.1921 If aiming at skin chromophore distribution mapping, the spectral information contained in smartphone digital images, taken under the built-in white LED illumination, appears to be insufficient. Narrowband filtering of the white illumination (e.g., by a set of interference filters)22 could improve the situation, but more promising from this point seems the use of specially designed small size multispectral add-on illuminators, e.g., based on narrowband color LEDs.23

This paper describes implementation of the RGB single-snapshot approach for the mapping of three main skin chromophores using a smartphone with attached compact three-wavelength laser illuminator. Image processing principles, design of the system, and the first results of its clinical tests are discussed below.

2.

Method

2.1.

Concept

The general concept of snapshot skin chromophore mapping at fixed wavelengths is illustrated in Fig. 1. Let us suppose that RGB color image of a skin pathology spot, surrounded by healthy skin, is captured under illumination that comprises only three equal intensity spectral lines at wavelengths λ1, λ2, and λ3 (the vertical lines in Fig. 1). With respect to the spectral sensitivity of an RGB image sensor and the cross-talk between its detection bands at the particular wavelengths, three monochromatic spectral images are extracted from the color image data set by the technique described in Refs. 1415.16.17. The three-chromophore skin model26 suggests that the dominant absorbers in skin at any of the fixed wavelengths λj (j=1, 2, 3) are oxyhemoglobin (further abbreviated as a), deoxyhemoglobin (b), and melanin (c)—see the crossings of their absorption curves with the vertical lines in Fig. 1. If the skin surface reflection is suppressed (e.g., by means of crossed polarizers), variations in chromophore composition would lead to changes of the diffusely reflected light intensities at each of the fixed wavelengths. Such variations in the pathology region relative to the healthy skin can be estimated by measuring reflected light intensities from equally sized regions of interest in the pathology (Ij) and the adjacent healthy skin (I0j). The ratios Ij/I0j at each pixel or pixel’s group of three monochromatic spectral images contain information on the concentration increase or decrease of all three regarded chromophores Δci (i=a, b, c), which can be further mapped over the whole image area.

Fig. 1

Absorption of three main skin chromophores at three fixed wavelengths.24,25

JBO_22_9_091508_f001.png

2.2.

Image Processing

The above-mentioned intensity changes at three exploited wavelengths due to absorption can be represented in terms of the Beer–Lambert–Bouguer law27

Eq. (1)

{ln(I1I01)=l1[Δca·ϵa(λ1)+Δcb·ϵb(λ1)+Δcc·ϵc(λ1)]ln(I2I02)=l2[Δca·ϵa(λ2)+Δcb·ϵb(λ2)+Δcc·ϵc(λ2)]ln(I3I03)=l3[Δca·ϵa(λ3)+Δcb·ϵb(λ3)+Δcc·ϵc(λ3)],
where ϵi(λj) is extinction coefficients of three regarded chromophores at three exploited wavelengths and lj is absorption path length in skin at a particular wavelength. Chromophore concentration increase or decrease at each image pixel (or selected group of pixels) is found by solving the linear equation system (Eq. 1) with abbreviated measured quantities kj=ln(Ij/I0j):

Eq. (2)

Δca=A1·k1+A2·k2+A3·k3Δcb=B1·k1+B2·k2+B3·k3Δcc=C1·k1+C2·k2+C3·k3.

The coefficients Aj, Bj, and Cj comprise numerical values of the corresponding chromophore extinction coefficients ϵi(λj) and absorption path lengths lj at the three exploited wavelengths. The respective extinction coefficients for oxyhemoglobin, deoxyhemoglobin, and melanin for our working wavelengths 448, 532, and 659 nm are presented in Table 1.

Table 1

The extinction coefficients of three main skin chromophores24,25 at three exploited wavelengths.

Chromophoreϵ (448 nm) (1/mol.cm)ϵ (532 nm) (1/mol.cm)ϵ (659 nm) (1/mol.cm)
Melanin19351170541
Oxyhemoglobin67,04445,400320
Deoxyhemoglobin173,32041,3363227

The Beer–Lambert–Bouguer law is originally addressed to absorption of a light beam passing through a slab of substance with thickness x. If this law is adapted to the diffusely remitted light geometry, slab thickness x has to be replaced by the mean path length lj of the backscattered photons inside the skin at a particular wavelength. To estimate the lj values with respect to light scattering properties, Monte Carlo (MC) simulations in a frame of the two-layered skin model28 (epidermal and dermal layers followed by the infinite subcutaneous tissue layer) were performed. Optical parameters of the tissue layers for the considered wavelengths were averaged from the number of available sources for typical Caucasian skin.2933 See Table 2.

Table 2

Optical properties of biotissue layers used in the simulations (λ, wavelength, μs, scattering coefficient, μa, absorption coefficient, g, anisotropy factor, n, refractive index).

λ (nm)μs (mm−1)μa (mm−1)gn
Epidermis659200.200.901.36
532270.600.851.36
448331.00.841.36
Dermis659180.120.901.40
532220.300.851.40
448270.650.841.40
Subcutaneous tissues65970.070.801.40
532100.400.761.40
448140.700.751.40

The simulations have been performed for various thicknesses of the epidermal layer (200, 150, and 100  μm) and a dermis of 1800  μm). In the simulations, uniform illumination beam with a diameter of 50 mm has been accounted. The diffusely backscattered (remitted) photons were collected from the same area accounting for the aperture angle of 10 deg. MC simulations were performed using a parallel (message passing interface) MC code for light propagation in biotissue-like media developed and verified in the earlier studies.3436 As the outcome, the distributions of the detected photons over their total path length and maximal penetration depth were generated. The mean path length and the mean penetration depth were calculated by weighted average taking into account the weight of the corresponding photon packets. The simulation results for skin with an epidermal thickness of 200  μm are presented in Fig. 2. MC simulations showed increased photon penetration depth and total path length with growing wavelength, the mean path length of the remitted photons being 4 to 5 times longer than their average penetration depth. The MC calculation results are presented in Table 3. One can see that the mean path length of photons in skin at a fixed wavelength is quite insensitive (<10%) to the twofold increase/decrease of the epidermal thickness. Consequently, different skin epidermal thicknesses should not seriously influence the data obtained by the proposed technique. The clinical images presented in Sec. 3 were processed using the MC-estimated mean path length values for 0.2-mm-thin skin epidermis.

Fig. 2

Monte Carlo simulated distributions of the detected photons over their total path length (a) and penetration depth (b) inside the skin (thickness of epidermis: 200  μm) for three exploited wavelengths. The mean path length and the mean penetration depth were calculated by weighted average taking into account weight of the corresponding photon packets.

JBO_22_9_091508_f002.png

Table 3

MC-estimated penetration depths and path length in skin of the remitted photons. Last column: ratio between the mean path length and the mean penetration depth.

Wavelength (nm)Epidermal thickness (mm)Mean penetration depth (mm)Mean path length in skin (mm)The calculated ratio factor
4480.200.160.835.19
0.150.170.885.18
0.100.180.925.11
5320.200.291.505.17
0.150.311.575.06
0.100.321.645.13
6590.200.752.933.91
0.150.762.953.88
0.100.762.983.92

To summarize, our algorithm for image processing includes the following steps:

  • 1. RGB image comprising pathology and adjacent healthy skin is captured under illumination by three discrete wavelengths (disabling all automatic settings of the camera).

  • 2. Three monochromatic spectral images are extracted from the digital image data set using the RGB crosstalk correction algorithm,14 with respect to actual relative intensities of the illumination lines.

  • 3. Equal area regions of interest are selected in the pathology and healthy skin zones of the monochromatic images.

  • 4. Pathology-to-healthy relative spectral reflectances Ij/I0j at the three exploited wavelengths and the values kj=ln(Ij/I0j) are calculated for all image pixels or selected pixel’s groups.

  • 5. The concentration increase/decrease of three chromophores in the pathology is calculated (2) for each pixel or pixel’s group.

  • 6. Distribution maps of three skin chromophores are constructed and displayed.

2.3.

Design of the Illumination Unit

A compact smartphone-compatible three-wavelength illuminator has been designed, assembled, and tested in laboratory and clinics. The illumination unit comprised six laser modules, an optical element for laser beam management, and an electronics compartment. Figure 3(a) shows the design details and outlook of the operating prototype with a smartphone on it [Fig. 3(b)]. To ensure uniform three-wavelength illumination of around target area, an optical element with a flat ring-shaped laser diffuser was developed.37 The illumination wavelengths 448, 532, and 659 nm were emitted by three pairs of compact 20-mW power laser modules (models PGL-DF-450 nm-20 mW-15011564, PGL-VI-1-532 nm-20 mW-15030443, and PGL-DF-655 nm-20 mW-150302232, Changchun New Industries Optoelectronics Tech. Co., Ltd). Laser modules [1, Fig. 3(a)—shown 3 out of 6] of each equal-wavelength pair were mounted at opposite sides on the internal wall of a hollow three-dimensional (3-D)-printed plastic shielding cylinder (2); the round bottom opening of this cylinder (diameter 40 mm) was in contact with skin and formed the field of view for the smartphone camera, situated 80 mm apart. All six coaxial laser beams were pointed to the 45-deg conical reflecting edge of a transparent disc (3) (beam collector) made of standard Plexiglas; after reflections they were turned radially toward the internal ring-shaped flat milky-Plexiglas diffuser (4). The upper and side surfaces of the collector/diffuser unit were mirrored by a vacuum-sputtered Al-coating. As a result, the flat diffuser (4) evenly illuminated the 65-mm distant skin target area simultaneously by three laser wavelengths. More details on the laser beam collector-diffuser are provided in Fig. 4.

Fig. 3

(a) 3-D model of the three-wavelength illumination unit and (b) the mobile prototype with smartphone on it. (1) Laser modules (three pairs, 448–532–659 nm), (2) shielding cylinder, (3) collector of laser beams, (4) flat ring-shaped diffuser of laser light, (5) sticky platform for the smartphone, (6) electronics compartment.

JBO_22_9_091508_f003.png

Fig. 4

Optical scheme of the laser beam collecting-diffusing element.36 The arrow denotes one of six coaxial laser beams that are turned radial by 45-deg conical reflecting edge of the transparent disk 3 and then downscattered by means of the internal diffusing ring 4. Upper surfaces of both disks are Al-mirrored. Dimensions: external diameter 50 mm, inner-outer diameters of the diffusing ring 10 to 40 mm, thickness 3 mm.

JBO_22_9_091508_f004.png

The smartphone—model Google Nexus5, 8 Mpx image sensor SONY IMX179 with known RGB-sensitivities38—was placed on a flat sticky platform (5) [Fig. 3(a)] with a round window for the smartphone rear camera, coaligned with the internal opening of the diffuser (4). The advantage of flat sticky platform design23,37,39 is the ability to fix most models of smartphones or tablet computers, independent of their dimensions; it was implemented by means of a double-sided tape. The round camera window was covered by a film polarizer (high contrast linear polarizing film40); another film with orthogonal direction of polarization covered the diffuser (4) from the bottom, so avoiding detection of skin surface-reflected light by the smartphone camera. The estimated spatial resolution of the imaging system was better than 0.1 mm.

A block diagram of the three-wavelength illumination device is shown in Fig. 5. Microcontroller (STM32F103) managed operation of the three-laser power modules via control of the laser power supply unit (PSU). Each pair of laser modules could be switched on and off independently by means of the control buttons. Two rechargeable batteries (Samsung 18650 3.7V Li-Ion, 2.8 Ah) and laser power supply circuits were placed in the 3-D-printed plastic compartment (6) below the platform (5). Target illumination intensity was adjusted to ensure linearity of photodetection by the smartphone camera; its automatic settings were switched off by the AZ Camera software, so the R-, G-, and B-outputs of the image sensor responded according to the spectral sensitivities of its three detection bands.38

Fig. 5

Block diagram of the three-wavelength illumination unit, where PSU stands for power supply unit.

JBO_22_9_091508_f005.png

The above-described optical design ensured acceptable uniformity of target illumination. It was initially checked at all combinations of illumination wavelengths (448, 532, and 659 nm) and image sensor detection channels (R, G, B), using white paper as the target. Figure 6 shows the obtained illumination intensity distributions, detected at separated RGB output channels in two cases: (i) when only one wavelength was used for illumination (659 nm detected at R-channel, 532 nm detected at G-channel, and 448 nm detected at B-channel, the upper row) and (ii) when all three wavelength of illumination were simultaneously switched on (the lower row).

Fig. 6

Uniformity of white paper illumination at different illumination–detection combinations: (a) R-image at single-wavelength 659-nm illumination; (b) G-image at single-wavelength 532-nm illumination; (c) B-image at single-wavelength 448-nm illumination; (d–f) R-, G-, and B-images at simultaneous three-wavelength illumination. Color scale represents relative intensities. Diameter of the illuminated area is 40 mm.

JBO_22_9_091508_f006.png

2.4.

Protocol of the Clinical Measurements

The mobile smartphone-illuminator system passed preliminary clinical case study under approval of the local ethics committee with written consent of white skin (phototypes 1 and 2) patients, 22- to 64-years old. Nine vascular and pigmented skin lesions were examined, including three pigmented nevi, three seborrheic keratosis, and three hemangiomas; all mentioned skin malformations were diagnosed by an experienced dermatologist. The main goal of clinical measurements was to check if the above-described hardware and software can provide physiologically feasible information on already diagnosed skin malformations in terms of increased/decreased contents of the three regarded chromophores. Statistical quantitative analysis of skin malformations at this stage of development was not planned.

The patient measurement protocol included the following steps:

  • (1) Standard dermatology procedure to diagnose the skin pathology;

  • (2) Brief introduction to the patient about the new technology and the device;

  • (3) Written consent of patient to participate;

  • (4) Capture of calibration image of gray color reference under three-wavelength illumination;

  • (5) Adjustment of the device on the examined skin pathology and capturing its image under three-wavelength illumination;

  • (6) Repeated capture of the gray color reference image under three-wavelength illumination.

The recorded pathology images were further processed by PC using the above-described algorithm. Reference images taken before and after the pathology image were compared to make sure that illumination conditions during the procedure have not changed. Gray reference (instead of white) was selected to avoid overexposure of the image sensor.

3.

Results

After processing the clinical images, skin chromophore maps were constructed and changes of malformation’s chromophore content with respect to the adjacent healthy skin could be evaluated. Clinical models1 assume that the pigmented skin lesions represent an abnormal increase of epidermal melanin and the vascular lesions—increased supply of dermal arterial blood (rich of oxyhemoglobin) to the superficial layers of skin. As initially expected, we observed notable melanin content increase in all cases of both pigmented malformations: nevi and seborrheic keratosis. Increase of oxyhemoglobin content and decrease of deoxyhemoglobin were observed in all examined vascular malformations—hemangiomas. For illustration, images for these clinical cases related to each of the examined three pathologies are presented in Figs. 7Fig. 89.

Fig. 7

(a) RGB image (scale bar: 5 mm) and the corresponding maps representing chromophore concentration changes for three cases of pigmented nevi: (b) oxyhemoglobin, (c) deoxyhemoglobin, and (d) melanin. Units of the color scale: millimoles (mM).

JBO_22_9_091508_f007.png

Fig. 8

(a) RGB image (scale bar: 5 mm) and the corresponding maps representing chromophore concentration changes for three cases of pigmented seborrheic keratosis: (b) oxyhemoglobin, (c) deoxyhemoglobin, (d) melanin. Units of the color scale: millimoles (mM).

JBO_22_9_091508_f008.png

Fig. 9

(a) RGB image (scale bar: 5 mm) and the corresponding maps representing chromophore concentration changes for three cases of vascular hemangioma: (b) oxyhemoglobin, (c) deoxyhemoglobin, (d) melanin. Units of the color scale: millimoles (mM).

JBO_22_9_091508_f009.png

The color photo of a malformation, taken under simultaneous three-wavelength illumination (a), is compared with the obtained concentration distribution maps of oxyhemoglobin (b), deoxyhemoglobin (c), and melanin (d). All pathology data are related to the adjacent healthy skin (Δc=0) so that only increased and decreased concentrations are quantified by colors. Figure 7 shows how melanin content increases in nevi, without essential changes in the hemoglobin content. Similar responses were obtained from the other pigmented skin malformation, seborrheic keratosis (Fig. 8). Quite different chromophore composition changes were recorded for vascular hemangiomas (Fig. 9); melanin concentration in the pathology remained practically unchanged while the oxyhemoglobin content notably increased and the deoxyhemoglobin content decreased in comparison to healthy skin.

The color scales in Figs. 79 are calibrated in millimoles, so the increased or decreased chromophore content can be quantitatively analyzed and/or compared. It might help dermatologists to characterize the severity of pathology if clinical thresholds for particular chromophores in the malformation are established.

The grainy structure of chromophore maps at Figs. 79 is most probably caused by formation of laser speckles on the skin surface. This is a coherent effect and cannot be avoided even in scattered laser light.

4.

Discussion

Generally, the obtained results have confirmed viability of the proposed approach for single-snapshot skin chromophore mapping with a smartphone. The compact add-on illuminator design [Figs. 3(a) and 4] ensured uniform skin illumination simultaneously at three-laser wavelengths (Fig. 6). The previously developed RGB monochromatic imaging concept, complemented by MC simulations of photon trajectories in skin, was successfully applied for mapping of pathologic concentration changes of skin melanin, oxyhemoglobin, and deoxyhemoglobin (Figs. 79). Results of this study may facilitate further developments of relatively simple and compact optical equipment for clinical diagnostics and monitoring; they also highlight the directions of further research, related to yet unsolved problems.

One of them is insufficient quality of chromophore maps due to laser speckles formed on the illuminated skin. In order to obtain speckle-free images, incoherent spectral line sources could be used. As an option, we assembled a three-wavelength illuminator based on high frequency electrodeless spectral lamps filled with an Hg–Cd vapor mixture. However, such an illuminator required an additional high quality multibandpass filter which made the design too expensive, so probably more efforts are needed to develop other kinds of speckle-free multiwavelength illumination sources.

Another issue of further improvements is the skin absorption model; only three main skin chromophores were considered in this study, while bilirubin, β-carotene, lipids, and water may also contribute as skin absorbers.3,29 Therefore, image-processing algorithms based on more advanced skin models (e.g., 4-chromophore or 5-chromophore) are to be developed in the future, along with further hardware developments. Smartphones comprise the conventional RGB image sensors, so only three output channels (R-, G-, and B-images) can be exploited; it limits to three the number of chromophores to be mapped by one snapshot. In this respect, a technology using several sets of simultaneous three-wavelength illumination should be developed in the future.

From the point of clinicians, the above-described technique would become useful for quantitative diagnostics only when a “gold standard” that sets the threshold (critical) values for increased chromophore concentrations, related to particular pathologies, will be available. To achieve that, serious clinical trials with a much larger number of patients and statistical analysis of the results have to be carried out. Severity of skin pathology could be quantified also in terms of chromophore concentration ratio “pathology/healthy”; further research in this direction would be necessary as well. In addition, results of the image processing—chromophore distribution maps—should appear on the screen of the used smartphone. Two approaches for that are to be examined: development of specific software application for particular (most popular) smartphone models or using a distant server that receives the captured images via mobile network and after processing sends back the calculated chromophore maps. The second option seems advantageous since it can be applied independently on the used smartphone model. As for potential hardware improvements, more handy for clinical use probably would be a separate multiwavelength illuminator which is first properly placed on the skin pathology and only then the smartphone attached for image capturing.

The mean path length in skin of the backscattered photons has been evaluated by means of MC simulations in this work. It would be useful to develop in the future techniques for direct determination of this parameter, e.g., using the photon time-of-flight measurement technique.41

5.

Conclusions

This study confirmed the possibility to map three main skin chromophores by a single snapshot of a smartphone camera under combined spectral line illumination. Millions of professionals nowadays use smartphones, so the proposed relatively simple and inexpensive approach potentially may find real clinical and forensic applications. The main challenges of the future work will be: (i) to minimize or fully avoid laser speckle artifacts in spectral images of the illuminated skin area; (ii) to improve the image-processing algorithms and hardware details; (iii) to expand the range of chromophores that can be mapped, e.g., to add bilirubin which shows up in cases of skin bruises and postsurgery recovery.

Disclosures

No conflicts of interest, financial or otherwise, are declared by the authors.

Acknowledgments

This study was supported by the Latvian National Research Program SOPHIS under the Grant agreement #10-4/VPP-4/11. Kind assistance of Dr. Aivars Vembris (ISSP, UL) in production of the light collecting/diffusing element is highly appreciated. AB acknowledges the support provided by Academy of Finland (Grant Nos. 296408 and 290596) and CSC–IT Center for Science, Finland, for computational resources. The fruitful collaboration between groups of two universities was facilitated by the European Commission COST action BM1205.

References

1. 

S. Chadwick, R. Heath and M. Shah, “Abnormal pigmentation within cutaneous scars: a complication of wound healing,” Indian J. Plast. Surg., 45 (2), 403 –411 (2012). http://dx.doi.org/10.4103/0970-0358.101328 Google Scholar

2. 

S. Cotton, “Astron Clinica SIAscope V,” (2017) http://www.panchromos.com/stories/astronclinica/ February ). 2017). Google Scholar

3. 

R. R. Anderson and J. A. Parrish, “The optics of human skin,” J. Invest. Dermatol., 77 (1), 13 –19 (1981). http://dx.doi.org/10.1111/1523-1747.ep12479191 JIDEAE 0022-202X Google Scholar

4. 

P. Valisou et al., “New closed-form approximation for skin chromophore mapping,” J. Biomed. Opt., 16 (4), 046012 (2011). http://dx.doi.org/10.1117/1.3562976 JBOPFO 1083-3668 Google Scholar

5. 

E. C. Ruvolo et al., “Use of spectral imaging for documentation of skin parameters in face lift procedure,” Proc. SPIE, 7548 75480A (2010). http://dx.doi.org/10.1117/12.848213 PSISDG 0277-786X Google Scholar

6. 

D. N. Stratis et al., “Comparison of acousto-optic and liquid crystal tunable filters for laser-induced breakdown spectroscopy,” Appl. Spectrosc., 55 (8), 999 –1004 (2001). http://dx.doi.org/10.1366/0003702011953144 Google Scholar

7. 

L. Randeberg, E. Larsen and L. Svaasand, “Characterization of vascular structures and skin bruises using hyperspectral imaging, image analysis and diffusion theory,” J. Biophoton., 3 (1–2), 53 –65 (2010). http://dx.doi.org/10.1002/jbio.200910059 Google Scholar

8. 

D. Yudovsky et al., “Assessing diabetic foot ulcer development risk with hyperspectral tissue oximetry,” J. Biomed. Opt., 16 (2), 026009 (2011). http://dx.doi.org/10.1117/1.3535592 JBOPFO 1083-3668 Google Scholar

9. 

F. Vasefi et al., “Separating melanin from hemodynamics in nevi using multimode hyperspectral dermoscopy and spatial frequency domain spectroscopy,” J. Biomed. Opt., 21 (11), 114001 (2016). http://dx.doi.org/10.1117/1.JBO.21.11.114001 JBOPFO 1083-3668 Google Scholar

10. 

I. Kuzmina et al., “Contact and contactless diffuse reflectance spectroscopy: potential for recovery monitoring of vascular lesions after intense pulsed light treatment,” J. Biomed. Opt., 16 (4), 040505 (2011). http://dx.doi.org/10.1117/1.3569119 JBOPFO 1083-3668 Google Scholar

11. 

C. H. Morimoto and R. Chellappa, “Evaluation of image stabilization algorithms,” IEEE Trans. Acoust. Speech Signal Process., 5 2789 –2792 (1988). http://dx.doi.org/10.1109/ICASSP.1998.678102 IETABA 0096-3518 Google Scholar

12. 

D. Jakovels and J. Spigulis, “RGB imaging system for mapping and monitoring of hemoglobin distribution in skin,” Proc. SPIE, 8158 8158OR (2011). http://dx.doi.org/10.1117/12.893789 PSISDG 0277-786X Google Scholar

13. 

D. Kapsokalyvas et al., “Spectral morphological analysis of skin lesions with a polarization multispectral dermoscope,” Opt. Express, 21 (4), 4826 –4840 (2013). http://dx.doi.org/10.1364/OE.21.004826 OPEXFF 1094-4087 Google Scholar

14. 

J. Spigulis and L. Elste, “Method and device for imaging of spectral reflectance at several wavelength bands,” (2012). Google Scholar

15. 

J. Spigulis and L. Elste, “Single snapshot RGB multispectral imaging at fixed wavelengths: proof of concept,” Proc. SPIE, 8937 89370L (2014). http://dx.doi.org/10.1117/12.2039442 PSISDG 0277-786X Google Scholar

16. 

J. Spigulis, D. Jakovels and L. Elste, “Towards single snapshot multispectral skin assessment,” Proc. SPIE, 8216 82160L (2012). http://dx.doi.org/10.1117/12.908967 PSISDG 0277-786X Google Scholar

17. 

J. Spigulis and I. Oshina, “Snapshot RGB mapping of skin melanin and hemoglobin,” J. Biomed. Opt., 20 (5), 050503 (2015). http://dx.doi.org/10.1117/1.JBO.20.5.050503 JBOPFO 1083-3668 Google Scholar

18. 

Anonymous, “2 billion consumers worldwide to get smart(phones) by 2016,” (2016) http://www.emarketer.com/Article/2-Billion-Consumers-Worldwide-Smartphones-by-2016/1011694 November ). 2016). Google Scholar

19. 

A. C. Brewer et al., “Mobile applications in dermatology,” J. Am. Med. Assoc. Dermatol., 149 (11), 1300 –1304 (2013). http://dx.doi.org/10.1001/jamadermatol.2013.5517 Google Scholar

20. 

N. MacKinnon et al., “Melanoma detection using smartphone and multimode hyperspectral imaging,” Proc. SPIE, 9711 971117 (2016). http://dx.doi.org/10.1117/12.2222415 PSISDG 0277-786X Google Scholar

21. 

J. A. Wolf et al., “Diagnostic inaccuracy of smartphone applications for melanoma detection,” J. Am. Med. Assoc. Dermatol., 149 (4), 422 –426 (2013). http://dx.doi.org/10.1001/jamadermatol.2013.2382 Google Scholar

22. 

S. Kim et al., “Smartphone-based multispectral imaging: system development and potential for mobile skin diagnosis,” Biomed. Opt. Express, 7 (12), 5294 (2016). http://dx.doi.org/10.1364/BOE.7.005294 BOEICL 2156-7085 Google Scholar

23. 

I. Kuzmina et al., “Study of smartphone suitability for mapping of skin chromophores,” J. Biomed. Opt., 20 (9), 090503 (2015). http://dx.doi.org/10.1117/1.JBO.20.9.090503 JBOPFO 1083-3668 Google Scholar

24. 

S. Prahl, “Tabulated molar extinction coefficient for hemoglobin in water,” (2016) http://omlc.ogi.edu/spectra/hemoglobin/summary.html November ). 2016). Google Scholar

25. 

T. Sarna and H. M. Swartz, “The physical properties of melanin,” (2016) http://omlc.ogi.edu/spectra/melanin/eumelanin.html November ). 2016). Google Scholar

26. 

D. Jakovels and J. Spigulis, “2-D mapping of skin chromophores in the spectral range 500–700 nm,” J. Biophoton., 3 (3), 125 –129 (2010). http://dx.doi.org/10.1002/jbio.v3:3 Google Scholar

27. 

B. M. Tissue, “Beer-Lambert law,” (2016) http://elchem.kaist.ac.kr/vt/chem-ed/spec/beerslaw.htm November ). 2016). Google Scholar

28. 

D. Yudovsky and L. Pilon, “Retrieving skin properties from in vivo spectral reflectance measurements,” J. Biophoton., 4 (5), 305 –314 (2011). http://dx.doi.org/10.1002/jbio.v4.5 Google Scholar

29. 

S. Jacques, “Optical properties of biological tissues: a review,” Phys. Med. Biol., 58 R37 –R61 (2013). http://dx.doi.org/10.1088/0031-9155/58/11/R37 PHMBA7 0031-9155 Google Scholar

30. 

G. Petrov et al., “Human tissue color as viewed in high dynamic range optical spectral transmission measurements,” Biomed. Opt. Express, 3 (9), 2154 –2161 (2012). http://dx.doi.org/10.1364/BOE.3.002154 BOEICL 2156-7085 Google Scholar

31. 

S.-H. Tseng, A. Grant and A. Durkin, “In vivo determination of skin near-infrared opticalproperties using diffuse optical spectroscopy,” J. Biomed. Opt., 13 (1), 014016 (2008). http://dx.doi.org/10.1117/1.2829772 JBOPFO 1083-3668 Google Scholar

32. 

M. Kirillin et al., “Simulation of optical coherence tomography images by Monte Carlo modeling based on polarization vector approach,” Opt. Express, 18 (21), 21714 –21724 (2010). http://dx.doi.org/10.1364/OE.18.021714 OPEXFF 1094-4087 Google Scholar

33. 

A. Bashkatov et al., “Optical properties of human skin, subcutaneous and mucous tissues in the wavelength range from 400 to 2000 nm,” J. Phys. D Appl. Phys., 38 2543 –2555 (2005). http://dx.doi.org/10.1088/0022-3727/38/15/004 Google Scholar

34. 

A. Bykov, A. Doronin, I. Meglinski, “Light propagation in turbid tissue-like scattering media,” Deep Tissue Imaging with Linear and Non-linear Optics, Pan Stanford Publishing Pte Ltd., Singapore (2017). Google Scholar

35. 

A. V. Bykov, A. V. Priezzhev and R. Myllylä, “Visualisation of structural inhomogeneities in strongly scattering media using the method of spatially-resolved reflectometry: Monte Carlo simulation,” Quant. Electron., 41 (6), 557 –563 (2011). http://dx.doi.org/10.1070/QE2011v041n06ABEH014642 QUELEZ 1063-7818 Google Scholar

36. 

L. P. Bass et al., “Parallel algorithm of the discrete ordinate method to laser impulse propagation simulation in turbid media,” Nuovo Cimento C, 33 (1), 39 –46 (2010). Google Scholar

37. 

J. Spigulis et al., “Method and device for smartphone mapping of tissue compounds,” WO2017/012675A1, Riga, Latvia (2017).

38. 

T. Nomura, , Sony LSI Design Incorporated, private correspondence.T. Nomura, , Sony LSI Design Incorporated, private correspondence.

39. 

A. Lihachev et al., “Autofluorescence imaging of basal cell carcinoma by smartphone RGB camera,” J. Biomed. Opt., 20 (12), 120502 (2015). http://dx.doi.org/10.1117/1.JBO.20.12.120502 JBOPFO 1083-3668 Google Scholar

40. 

Anonymous, “High contrast linear polarizing film,” (2017) https://www.edmundoptics.com/optics/polarizers/linear-polarizers/high-contrast-linear-polarizing-film/3435/ February ). 2017). Google Scholar

41. 

A. Dzerve et al., “Time of flight for photons in human skin,” J. Biomed. Photonics Eng., 2 (3), 030301 (2016). http://dx.doi.org/10.18287/JBPE16.02.030301 Google Scholar

Biography

Janis Spigulis is the head of Biophotonics Laboratory at the Institute of Atomic Physics and Spectroscopy (IAPS) and professor at the Physics Department, University of Latvia. His research is focused on optical noncontact human skin diagnostics and monitoring.

Ilze Oshina is a research assistant at Biophotonics Laboratory of IAPS and a postgraduate student at the Faculty of Physics and Mathematics, University of Latvia. Her research interests are related to in vivo skin imaging and chromophore mapping by advanced photonic technologies. She had participated in several research projects related to skin diagnostics.

Anna Berzina is a research assistant at Biophotonics Laboratory of IAPS and certified dermatologist, expert in noninvasive diagnostics of skin and its treatment with lasers and other light-emitting devices. She has taken part in research projects related to optical skin diagnostics and recovery monitoring and is coauthor of more than 10 scientific publications.

Alexander Bykov received his MSc degree in 2005 and his PhD in 2008 from Lomonosov Moscow State University, Russia. He received his DSc (Tech.) degree from the Faculty of Technology, University of Oulu, Finland, in 2010. Currently, he is a senior researcher in the Optoelectronics and Measurement Techniques Laboratory, University of Oulu, Finland.

© 2017 Society of Photo-Optical Instrumentation Engineers (SPIE) 1083-3668/2017/$25.00 © 2017 SPIE
Janis Spigulis, Ilze Oshina, Anna Berzina, and Alexander Bykov "Smartphone snapshot mapping of skin chromophores under triple-wavelength laser illumination," Journal of Biomedical Optics 22(9), 091508 (2 March 2017). https://doi.org/10.1117/1.JBO.22.9.091508
Received: 30 November 2016; Accepted: 10 February 2017; Published: 2 March 2017