31 July 2023 Spectral correction of light emitting diodes enables accurate hydration ratio calculation using narrowband diffuse reflectance spectroscopy
Jesse H. Lam, Jeonghun Kim, Kelsey J. Tu, Dongbin Kim, Sehwan Kim
Author Affiliations +
Abstract

Significance

Light emitting diodes (LEDs) are commonly utilized for tissue spectroscopy due to their small size, low cost, and simplicity. However, LEDs are often approximated as single-wavelength devices despite having relatively broad spectral bandwidths. When paired with photodiodes, the wavelength information of detected light cannot be resolved. This can result in errors during chromophore concentration calculations. These errors are particularly apparent when analyzing water and fat in the 900 to 1000 nm window where the spectral bandwidth of LEDs can encompass much of the analysis region, resulting in intense crosstalk.

Aim

We utilize and present a spectral correction (SC) algorithm to correct for the spectral bandwidth of LEDs. We show the efficacy using a narrowband technique of spectrally broad and overlapping LEDs.

Approach

Narrowband diffuse reflectance spectroscopy (nb-DRS), a technique capable of quantifying the hydration ratio (RH2O) of turbid media, was utilized. nb-DRS typically requires a broadband light source and spectrometer. We reduce the hardware to just five LEDs and a photodiode detector, relying on SC to compensate for spectral crosstalk. The effectiveness of our SC approach was tested in simulations as well as in an emulsion phantom and limited selection of human tissue.

Results

In simulations, we show that calculated RH2O errors increased with the spectral bandwidth of LEDs but could be corrected using SC. Likewise, in emulsions, we found an average error of 8.7% (maximum error 14%) if SC was not used. By contrast, applying SC reduced the average error to 2.2% (maximum error of 6.4%). We show that despite utilizing multiple, spectrally broad, and overlapping LEDs, SC was still able to restore the performance of our narrowband method, making it comparable to a much larger full broadband system.

1.

Introduction

Maintaining hydration is necessary for human survival, yet it remains a challenge to monitor continuously and portably by optical methods. Dehydration can result in severe athletic and cognitive deterioration under 10% body water loss, and potential death if it further progresses.1,2 Tissue hydration was also found to be an important biomarker in other fields, such as breast cancer research,35 dermatology,6,7 and critical care medicine.8,9 Approaches for estimating water content have been previously validated using near-infrared (NIR) optical techniques, such as diffuse optical spectroscopy1012 and spatial frequency domain imaging (SFDI).13,14

In our previous work, we demonstrated that narrowband diffuse reflectance spectroscopy (nb-DRS) can calculate the hydration ratio (RH2O), which tracked closely with absolute water fraction. This was achieved by analysis of diffuse reflectance in the far NIR 900 to 1000 nm wavelength region15 using a broadband lamp and spectrometer. We defined RH2O as a ratio of water (H2O) and lipid (FAT):

Eq. (1)

RH2O(%)=100×H2OH2O+FAT.

Due to the low hardware requirements of nb-DRS, we theorized that it could be a suitable candidate for miniaturization as a wearable hydration monitor. For the light source, multiple wavelength light emitting diodes (LEDs) could be utilized in lieu of the broadband lamp as a smaller, lower power, and low-cost substitute.16

However, for detection, transitioning from spectrometers to photodiodes has proved to be challenging. One drawback is the loss of high-resolution spectral data as photodiodes cannot resolve wavelength information. This limitation is exacerbated by LEDs having a non-negligible spectral full-width at half-maximum (FWHM), potentially leading to spectral crosstalk if not corrected.1719 For some spectroscopic techniques, LEDs can be approximated using the peak wavelength. Using this approach, Mazhar et al.20 reported a <5% error for SFDI when the FWHM of LEDs were below 30 nm. Kurata et al.21,22 proposed averaging the chromophore extinction coefficients based on the bandwidth of the sources’ LEDs. However, this approach is not compatible with nb-DRS as a narrowband (900 to 1000 nm) technique. Utilizing the average extinction coefficients of given broad (>30  nm) LEDs is not feasible as the average would encompass too much of our analysis region including both H2O and FAT absorption peaks. Other groups have adopted an integration approach that directly uses the LED emission spectra during calculations.2326 This method requires careful characterization of the system LEDs and is more appropriate for nb-DRS.

We utilize and present a specific integration algorithm for spectral correction (SC) that can mitigate crosstalk errors for narrowband techniques even when using multiple, spectrally broad, and overlapping LEDs. We show the effectiveness of this algorithm in simulations and an emulsion phantom spanning a wide range of RH2O values. Measurements of human abdomen and thenar tissue are also performed. Pseudo-code and MATLAB code for our application of SC for nb-DRS are also provided. We hope that our efforts will facilitate the development of tissue hydration monitors using nb-DRS and encourage application of SC to other modalities as well.

2.

Materials and Methods

2.1.

Optical Instrumentation

We constructed an nb-DRS probe utilizing two approaches to the technique. To serve as a reference device, a full broadband nb-DRS method was used, as previously reported.15 A tungsten halogen lamp (HL-2000-HP, Ocean Insight) was coupled using an optical fiber with a core diameter of 400  μm. For detection, a spectrometer (ULS2048XL, Avantes B.V.), which was fiber coupled using a 1000  μm diameter optical fiber, was utilized.

As for the other nb-DRS approach, we utilized a customized multi-package LED containing five LEDs (Shenzhen Best LED Opto-electronic Co., Ltd.) with spectra shown in Fig. 1(a). The peak wavelengths of the LEDs were 907, 929, 944, 949, and 976 nm, whereas the FWHM were 29, 38, 56, 35, and 53 nm, respectively. Compared with typical absorption spectra for tissue containing oxyhemoglobin (HbO2), deoxyhemoglobin (HbR), H2O, and FAT chromophores, e.g., human breast tissue27 shown in Fig. 1(b), it can be seen that our LED spectra were quite broad. In particular, the 944, 949, and 976 nm LED spectra significantly overlapped with both the FAT and H2O absorption peaks. There was also significant overlap between LEDs. A photodiode (S121C, Thorlabs, Inc.), which was fiber coupled by a 1000  μm optical fiber, was used for detection. The LED and broadband approaches were mounted in a cross-configuration.

Fig. 1

Spectra of the LEDs in the device. Vertical dashed-dot lines indicate the wavelength analysis region for nb-DRS. (a) Spectra for the LEDs in the nb-DRS instrument. LEDs are labeled by their peak wavelengths, shown in the legend. (b) Chromophore absorption spectra for human breast tissue containing 12.6  μM HbO2, 27.9  μM HbR, 45.1% H2O, and 29.8% FAT.2729 It can be seen that the spectral bandwidth of several LEDs were wide enough to overlap each other as well as both the H2O and FAT absorption peaks in this wavelength range.

JBO_28_7_075005_f001.png

Optical components were mounted in a single 3D-printed probe with a source to detector distance of 15 mm. Due to the slight spatial distribution of the LEDs in the package, the actual source–detector separation for each individual LED varied between 13.5 and 15.5 mm, as measured using a caliper. A 0.6 mm thick transparent plastic sheet was placed between the probe and emulsion to protect the optical components. Possible effects of light piping due to the plastic sheet were empirically found to be negligible, with tissue RH2O calculations only differing by 1%. Device calibration and LED spectra were measured using a reflectance standard (SRS-02-020, Labsphere Inc.). Using the measurement from the reflectance standard as a reference, instrument characteristics, such as different LED powers for each wavelength and photodiode sensitivity, can be approximated. However, we note that, without a direct measurement of the photodiode responsivity, the detector wavelength sensitivity function cannot be perfectly corrected and may contribute to experimental error.

The LEDs were powered and controlled using an Arduino Nano. The broadband lamp, spectrometer, and photodiode were operated using custom MATLAB 2021b (The MathWorks, Inc.) code.

2.2.

Simulations

A set of simulated broadband reflectance spectra were generated using the steady-state diffusion model30 with varying RH2O values and a constant reduced scattering of 1  mm1. LED profiles with center wavelengths equal to the ones in our device were simulated as Gaussian curves with FWHM values increasing from 10 to 60 nm. The simulated broadband reflectance was then transformed into simulated LED reflectance using the Gaussian curves.

We then attempted to calculate RH2O by two methods. First, we used the original nb-DRS algorithm, which does not account for the spectrum of the light source. Instead, the LEDs were assumed to be represented by the peak wavelength. Second, we used our SC algorithm, which assumes data from the photodiode to be an integration of reflectance over wavelength based on the measured LED spectra.

2.3.

Sample Measurements

An emulsion phantom was constructed following processes similar to those described by Ohmae et al.,11 with slight modifications to maintain the phantom in liquid form instead of a solid gelatin. An oil and water emulsion phantom was fabricated with an initial oil to water ratio of 90 to 10 using soy lecithin (2% by weight) as the emulsifier. The solutions were then mixed using a high-speed vacuum blender (i8800, Jiaxiang Electric Co., Ltd.). Next, the phantom was diluted by adding 100 mL of water for each step. By maintaining water as the continuous medium, the phantom could be diluted without separation of the constituents.31 This approach also allowed for measurements across a large range of phantom RH2O values without needing to construct a new phantom for each step. After 15 dilution steps, the final expected RH2O was 74.3%. After every step, the phantom was homogenized using a hand-held mixer. The probe was also lifted, cleaned, and then replaced such that the face of the probe rested on the liquid surface and was not submerged.

Finally, measurements were performed on two healthy human subjects (31 year old male and 30 year old female). Abdomen and thenar sites were selected as representative examples of fatty and lean tissue, respectively. The source–detector separation for the LED and photodiode combination was shortened to 11 mm to improve the signal-to-noise ratio. The source–detector separation for the broadband nb-DRS optical fibers remained unchanged. Although not a significant factor for homogeneous samples, such as emulsions, the differences in source–detector separation and optical geometry could result in slightly different measurement volumes in tissue. Hemoglobin was included during chromophore fits.

2.4.

Data Processing

Data processing was done by custom code written in MATLAB. Our SC algorithm was employed to account for the spectral bandwidth of our LEDs, with the pseudo-code shown in Table 1. First, calibration was performed using a reflectance standard. In step 2, the measured LED profile data were normalized with respect to the maximum intensity for each wavelength to be used as weighting in a later step. Next, broadband total absorption was calculated using known chromophore extinction coefficient tables and an initial chromophore concentration guess (e.g., 50% H2O and 50% FAT). Similar to our previous work, we assumed a flat, constant scattering of 1.0  mm1. Due to the power law of scattering, this was found to be a reasonable assumption in the far NIR of 900 to 1000 nm. Furthermore, as shown in our previous work, this scattering assumption has little effect on calculated RH2O.15 Broadband reflectance was then calculated using absorption and reduced scattering values. The high-spectral resolution broadband reflectance data were transformed into low-resolution simulated photodiode reflectance (e.g., 1 data point per LED) by multiplication with the LED profile weights and then summed over wavelength. The simulated photodiode reflectance was then minimized by comparison with the actual measured data with RH2O as the final output.

Table 1

Pseudo-code detailing SC used for our nb-DRS system.

Pseudo-code: SC algorithm for nb-DRS
INPUT:
D are measurement data from the photodiode for each LED
C are calibration data for the photodiode for each LED
S are the LED spectral measurements
ε are the chromophore extinction coefficients for all wavelengths
λ are the wavelengths for extinction coefficients
N are the chromophore concentrations
f is the steady-state diffusion model function
θ are all other diffusion model inputs (e.g., index of refraction and scattering)
OUTPUT:
H is the calculated hydration ratio
START:
1. CALIBRATE measurement data D using calibration data C
  R=DC
2. NORMALIZE spectral information of the LEDs S
  Snorm=Smax(S)
3. FIT to solve for chromophore concentration N
  N=argminN|RRnorm(N;θ)|2
  WHERE:
   FOR j=1 to length of R
   Rsim(j)=λ[f(N,ε,λ,θ)×Snorm(λ,j)]
   END
   Rnorm=R(1)×RsiRsim(1)
4. RETURN final output H
  H=100×NH2ONH2O+NFAT

3.

Results and Discussion

3.1.

Simulations

A set of reflectance spectra were simulated using LEDs with increasing FWHM ranging from 10 to 60 nm. RH2O values were then recovered using nb-DRS with and without our SC algorithm, as shown in Fig. 2. When using LEDs with an FWHM of only 10 nm, the performance for either approach was similar (1% error). However, as the bandwidth of the LEDs increased, the calculations without SC became progressively erroneous. As the FWHM reached 60 nm, the average absolute error had ballooned to 13.4%. By contrast, calculations of RH2O with our SC algorithm maintained the same level of accuracy for all simulated FWHM values (within 1%).

Fig. 2

Simulated data to compare fitting approaches with and without our SC algorithm. The dashed line represents the line of identity. In the absence of SC, light from each LED was assumed to be the peak wavelength. When the FWHM of the LEDs was sufficiently narrow, this assumption was reasonable. However, as the spectral width of the LED increased, this assumption became invalid and the estimated RH2O was increasingly erroneous. By contrast, using our SC algorithm, an accurate RH2O was recovered despite increasing FWHM values.

JBO_28_7_075005_f002.png

3.2.

Emulsion Phantoms

Our SC algorithm was further tested using an emulsion phantom. Three methods were compared, as seen in Fig. 3. First, the reference system, which used a broadband light source and spectrometer detector, is shown. This method performed most accurately with an average and standard deviation error of 1.4%±0.7%. The maximum error was 2.2%.

Fig. 3

Emulsion RH2O measured by nb-DRS using a broadband light source, LED source without SC, and LED source with SC. The dashed line represents the line of identity. Note that the dilution volume added at each step was maintained at 100 mL, whereas the total phantom volume increased over time. This resulted in a non-linear dilution trend, with larger RH2O increases at the start of the experiment and smaller RH2O changes near the end.

JBO_28_7_075005_f003.png

Next, the LEDs were utilized in conjunction with a photodiode detector. Effects of spectral crosstalk can be recognized not only by accuracy but also by sensitivity. For example, an LED at 929 nm might be considered sensitive to FAT, as the peak wavelength is matched with the lipid absorption peak. However, in reality, both H2O and FAT can affect reflectance measurements using this LED due to the wide spectral bandwidth. This was evident as, without SC, the dynamic range was poor. This was observed in both our simulations and emulsion phantom experiments. The error on the emulsion was found to be 8.7%±4.4% with a maximum error of 14.1%. Finally, using the same dataset, our SC algorithm was applied. This improved the error to 2.2%±1.7% with a maximum error of 6.4%. It can also be seen that applying SC restored the sensitivity, with a slope comparable to the full broadband approach.

Some discrepancy can be seen between the broadband approach and the LED (with SC) approach. The broadband approach has a much greater spectral resolution with 51 data points, allowing for more robust chromophore fitting. By contrast, the LED approach only has one datapoint for each LED. Furthermore, the LEDs are light sources with spectral bandwidths ranging from 30 to 50 nm. This spectral crosstalk can be corrected using our algorithm, but this lack of information cannot be perfectly compensated. Thus there are some small discrepancies between the broadband and LED (with SC) RH2O results.

In addition, attenuation by scattering was most intense at the beginning of the experiment when the lipid droplet concentration was highest. Although the photodiode in use had a large active diameter of 9.7 mm, it was fiber-coupled to a 1 mm core diameter optical fiber to produce a more compact probe and allow for a shorter source-detector separation. Future iterations of this device will include custom circuitry with a direct-contact photodiode to improve the signal-to-noise ratio of our measurements.

3.3.

Tissue Measurements

Measurement results on human tissue are shown in Table 2. The broadband and LED (with SC) approaches were able to clearly differentiate between abdomen and thenar tissue for the two subjects. RH2O on abdomen tissue was low due to the significant presence of adipose, ranging from 18.9% to 37.8% for the broadband approach and 22.7% to 37.8% for the LED (with SC) approach. By contrast, thenar tissue was a much leaner tissue with 77% as reported by the broadband system. Using SC, the LED system reported a range between 69% and 80.7%. This discrepancy could be due to slightly different volume measurements due to the differences in source–detector separation (15 versus 11 mm). For either approach, the two tissue types were clearly delineated, with abdomen tissue having a much lower RH2O than thenar, as expected. The last approach, LED (no SC), failed to properly fit the data. Although the emulsion only contained H2O and FAT, the inclusion of hemoglobin factors in live tissue likely exacerbated spectral crosstalk errors. Although SC can improve the performance of the LED approach, the technique is still ultimately limited by the number of LED and lack of high-resolution spectral information. Complex tissue is more computationally challenging due to the presence of other factors, such as hemoglobin, as well as other unaccounted chromophores. Accuracy could potentially be improved by the addition of LEDs and inclusion of other chromophores, such as myoglobin and melanin.

Table 2

RH2O estimations on human abdomen and thenar tissue. The broadband device utilized a 15 mm source–detector distance, whereas the LED device utilized an 11 mm source–detector distance.

TissueBroadband (%)LED (with SC) (%)LED (no SC) (%)
Abdomen 137.837.858.2
Abdomen 218.922.753.6
Thenar 175.080.760.6
Thenar 278.769.059.7

4.

Conclusion

We reported an SC algorithm that was shown to increase the accuracy of RH2O measurements using nb-DRS despite the usage of overlapping, broad spectrum LEDs in a narrow wavelength region. This was achieved by leveraging the known spectra of the LEDs during chromophore calculations. We showed the efficacy of SC in simulations as well as an emulsion phantom and a limited selection of tissues using nb-DRS. Through this work, we hope to facilitate the simplification and miniaturization of spectroscopic technologies and encourage the usage of SC, especially for narrowband LED-based devices.

Disclosures

J.K. is a PhD student at Dankook University and an employee of MEDiThings. J.H.L. and S.K. disclose patents owned by Dankook University related to nb-DRS technology. This research, including data collection, processing, and manuscript drafting, was conducted without participation from commercial entities.

Code, Data, and Materials Availability

Application of the algorithm for nb-DRS is available for viewing at Code Ocean ( https://doi.org/10.24433/CO.7379307.v1). Referenced chromophore extinction coefficient tables were not republished here but can be found in the original cited sources.

Acknowledgments

This work was supported by the National Research Foundation of Korea (NRF) funded by MSIT and MOE (NRF-2020H1D3A1A0408095814, NRF-2022R1I1A3073688, and NRF-2023K1A4A3A02057280).

References

1. 

A. Sarvazyan, A. Tatarinov and N. Sarvazyan, “Ultrasonic assessment of tissue hydration status,” Ultrasonics, 43 (8), 661 –671 https://doi.org/10.1016/j.ultras.2005.03.005 ULTRA3 0041-624X (2005). Google Scholar

2. 

S. N. Cheuvront, R. Carter and M. N. Sawka, “Fluid balance and endurance exercise performance,” Curr. Sports Med. Rep., 2 (4), 202 –208 https://doi.org/10.1249/00149619-200308000-00006 (2003). Google Scholar

3. 

N. Shah et al., “Combined diffuse optical spectroscopy and contrast-enhanced magnetic resonance imaging for monitoring breast cancer neoadjuvant chemotherapy: a case study,” J. Biomed. Opt., 10 (5), 051503 https://doi.org/10.1117/1.2070147 JBOPFO 1083-3668 (2005). Google Scholar

4. 

A. Cerussi et al., “In vivo absorption, scattering, and physiologic properties of 58 malignant breast tumors determined by broadband diffuse optical spectroscopy,” J. Biomed. Opt., 11 (4), 044005 https://doi.org/10.1117/1.2337546 JBOPFO 1083-3668 (2006). Google Scholar

5. 

L. L. de Boer et al., “Fat/water ratios measured with diffuse reflectance spectroscopy to detect breast tumor boundaries,” Breast Cancer Res. Treat., 152 (3), 509 –518 https://doi.org/10.1007/s10549-015-3487-z BCTRD6 (2015). Google Scholar

6. 

H. Arimoto, M. Egawa and Y. Yamada, “Depth profile of diffuse reflectance near-infrared spectroscopy for measurement of water content in skin,” Ski. Res. Technol., 11 (1), 27 –35 https://doi.org/10.1111/j.1600-0846.2005.00093.x (2005). Google Scholar

7. 

M. Attas et al., “Visualization of cutaneous hemoglobin oxygenation and skin hydration using near-infrared spectroscopic imaging,” Ski. Res. Technol., 7 (4), 238 –245 https://doi.org/10.1034/j.1600-0846.2001.70406.x (2001). Google Scholar

8. 

E. H. Starling, “On the absorption of fluids from the connective tissue spaces,” J. Physiol., 19 (4), 312 –326 https://doi.org/10.1113/jphysiol.1896.sp000596 JPHYA7 0022-3751 (1896). Google Scholar

9. 

S. M. Adler and J. G. Verbalis, “Disorders of body water homeostasis in critical illness,” Endocrinol. Metab. Clin. North Am., 35 (4), 873 –894 https://doi.org/10.1016/j.ecl.2006.09.011 ECNAER 0889-8529 (2006). Google Scholar

10. 

S. Merritt et al., “Comparison of water and lipid content measurements using diffuse optical spectroscopy and MRI in emulsion phantoms,” Technol. Cancer Res. Treat., 2 (6), 563 –569 https://doi.org/10.1177/153303460300200608 (2003). Google Scholar

11. 

E. Ohmae et al., “Stable tissue-simulating phantoms with various water and lipid contents for diffuse optical spectroscopy,” Biomed. Opt. Express, 9 (11), 5792 https://doi.org/10.1364/BOE.9.005792 BOEICL 2156-7085 (2018). Google Scholar

12. 

R. Nachabé et al., “Estimation of lipid and water concentrations in scattering media with diffuse optical spectroscopy from 900 to 1600 nm,” J. Biomed. Opt., 15 (3), 037015 https://doi.org/10.1117/1.3454392 JBOPFO 1083-3668 (2010). Google Scholar

13. 

B. Song et al., “Quantitative spatial mapping of tissue water and lipid content using spatial frequency domain imaging in the 900- to 1000-nm wavelength region,” J. Biomed. Opt., 27 (10), 105005 https://doi.org/10.1117/1.JBO.27.10.105005 JBOPFO 1083-3668 (2022). Google Scholar

14. 

J. Q. Nguyen et al., “Spatial frequency domain imaging of burn wounds in a preclinical model of graded burn severity,” J. Biomed. Opt., 18 (6), 066010 https://doi.org/10.1117/1.JBO.18.6.066010 JBOPFO 1083-3668 (2013). Google Scholar

15. 

J. H. Lam, K. J. Tu and S. Kim, “Narrowband diffuse reflectance spectroscopy in the 900–1000 nm wavelength region to quantify water and lipid content of turbid media,” Biomed. Opt. Express, 12 (6), 3091 https://doi.org/10.1364/BOE.425451 BOEICL 2156-7085 (2021). Google Scholar

16. 

S. Fantini et al., “Quantitative determination of the absorption spectra of chromophores in strongly scattering media: a light-emitting-diode based technique,” Appl. Opt., 33 (22), 5204 https://doi.org/10.1364/AO.33.005204 APOPAI 0003-6935 (1994). Google Scholar

17. 

K. Uludag et al., “Cross talk in the Lambert–Beer calculation for near-infrared wavelengths estimated by Monte Carlo simulations,” J. Biomed. Opt., 7 (1), 51 https://doi.org/10.1117/1.1427048 JBOPFO 1083-3668 (2002). Google Scholar

18. 

S. Umeyama and T. Yamada, “New cross-talk measure of near-infrared spectroscopy and its application to wavelength combination optimization,” J. Biomed. Opt., 14 (3), 034017 https://doi.org/10.1117/1.3147402 JBOPFO 1083-3668 (2009). Google Scholar

19. 

G. Strangman, M. A. Franceschini and D. A. Boas, “Factors affecting the accuracy of near-infrared spectroscopy concentration calculations for focal changes in oxygenation parameters,” Neuroimage, 18 (4), 865 –879 https://doi.org/10.1016/S1053-8119(03)00021-1 NEIMEF 1053-8119 (2003). Google Scholar

20. 

A. Mazhar et al., “Wavelength optimization for rapid chromophore mapping using spatial frequency domain imaging,” J. Biomed. Opt., 15 (6), 061716 https://doi.org/10.1117/1.3523373 JBOPFO 1083-3668 (2010). Google Scholar

21. 

T. Kurata et al., “Correction method for influence of tissue scattering for sidestream dark-field oximetry using multicolor LEDs,” Opt. Rev., 23 (6), 955 –967 https://doi.org/10.1007/s10043-016-0278-7 1340-6000 (2016). Google Scholar

22. 

T. Kurata et al., “Impact of vessel diameter and bandwidth of illumination in sidestream dark-field oximetry,” Biomed. Opt. Express, 6 (5), 1616 https://doi.org/10.1364/BOE.6.001616 BOEICL 2156-7085 (2015). Google Scholar

23. 

I. Oshina and J. Spigulis, “Beer–Lambert law for optical tissue diagnostics: current state of the art and the main limitations,” J. Biomed. Opt., 26 (10), 100901 https://doi.org/10.1117/1.JBO.26.10.100901 JBOPFO 1083-3668 (2021). Google Scholar

24. 

M. Marois et al., “Optimal wavelength selection for optical spectroscopy of hemoglobin and water within a simulated light-scattering tissue,” J. Biomed. Opt., 23 (7), 071202 https://doi.org/10.1117/1.JBO.23.4.041202 JBOPFO 1083-3668 (2018). Google Scholar

25. 

R. Hashimoto et al., “Two-wavelength oximetry of tissue microcirculation based on sidestream dark-field imaging,” J. Biomed. Opt., 24 (3), 031013 https://doi.org/10.1117/1.JBO.24.3.031013 JBOPFO 1083-3668 (2018). Google Scholar

26. 

C. Wang et al., “Extendable, large-field multi-modal optical imaging system for measuring tissue hemodynamics,” Biomed. Opt. Express, 11 (5), 2339 https://doi.org/10.1364/BOE.386197 BOEICL 2156-7085 (2020). Google Scholar

27. 

A. E. Cerussi et al., “Sources of absorption and scattering contrast for near-infrared optical mammography,” Acad. Radiol., 8 (3), 211 –218 https://doi.org/10.1016/S1076-6332(03)80529-9 (2001). Google Scholar

28. 

R. L. P. van Veen et al., “Determination of visible near-IR absorption coefficients of mammalian fat using time- and spatially resolved diffuse reflectance and transmission spectroscopy,” J. Biomed. Opt., 10 (5), 054004 https://doi.org/10.1117/1.2085149 JBOPFO 1083-3668 (2005). Google Scholar

29. 

L. Kou, D. Labrie and P. Chylek, “Refractive indices of water and ice in the 065- to 25-μm spectral range,” Appl. Opt., 32 (19), 3531 https://doi.org/10.1364/AO.32.003531 APOPAI 0003-6935 (1993). Google Scholar

30. 

A. Kienle and M. S. Patterson, “Improved solutions of the steady-state and the time-resolved diffusion equations for reflectance from a semi-infinite turbid medium,” J. Opt. Soc. Am. A, 14 (1), 246 https://doi.org/10.1364/JOSAA.14.000246 JOAOD6 0740-3232 (1997). Google Scholar

31. 

P. Dietemann et al., “A colloidal description of tempera and oil paints, based on a case study of Arnold Böcklin’s painting Villa am Meer II (1865),” e-Preserv. Sci., 11 (January), 29 –46 (2014). Google Scholar

Biographies of the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Jesse H. Lam, Jeonghun Kim, Kelsey J. Tu, Dongbin Kim, and Sehwan Kim "Spectral correction of light emitting diodes enables accurate hydration ratio calculation using narrowband diffuse reflectance spectroscopy," Journal of Biomedical Optics 28(7), 075005 (31 July 2023). https://doi.org/10.1117/1.JBO.28.7.075005
Received: 19 March 2023; Accepted: 14 July 2023; Published: 31 July 2023