12 January 2017 New-generation small animal positron emission tomography system for molecular imaging
Shiva Abbaszadeh, Craig S. Levin
Author Affiliations +
Abstract
The next generation of discoveries in molecular imaging requires positron emission tomography (PET) systems with high spatial resolution and high sensitivity to visualize and quantify low concentrations of molecular probes. The goal of this work is to assemble and explore such a system. We use cadmium zinc telluride (CZT) to achieve high spatial resolution, three-dimensional interaction positioning, and excellent energy resolution. The CZT crystals are arranged in an edge-on configuration with a minimum gap of 50  μm in a four-sided panel geometry to achieve superior photon sensitivity. The developed CZT detectors and readout electronics were scaled up to complete significant portions of the final PET system. The steering electrode bias and the amplitude of the analog signals for time measurement were optimized to improve performance. The energy resolution (at 511 keV) over 468 channels is 7.43±1.02% full-width-at-half-maximum (FWHM). The spatial resolution is 0.76±0.1  mm FWHM. The time resolution of six CZT crystals in coincidence with six other CZT crystals is 37 ns. With high energy and spatial resolution and the relatively low random rate for small animal imaging, this system shows promise to be very useful for molecular imaging studies.
© 2017 Society of Photo-Optical Instrumentation Engineers (SPIE) 2329-4302/2017/$25.00 © 2017 SPIE
Shiva Abbaszadeh and Craig S. Levin "New-generation small animal positron emission tomography system for molecular imaging," Journal of Medical Imaging 4(1), 011008 (12 January 2017). https://doi.org/10.1117/1.JMI.4.1.011008
Received: 30 July 2016; Accepted: 27 December 2016; Published: 12 January 2017
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Crystals

Imaging systems

Positron emission tomography

Sensors

Spatial resolution

Electrodes

Molecular imaging

RELATED CONTENT

Positron Emission Tomography: Instrumentation Perspectives
Proceedings of SPIE (November 01 1982)
HyperSPECT a new system for pre clinical imaging in...
Proceedings of SPIE (March 13 2009)
Position-sensitive CZT detector module
Proceedings of SPIE (July 01 1998)
Two-bent-crystal schemes for monochromatic x-ray imaging
Proceedings of SPIE (September 18 1995)

Back to Top