7 November 2022 Comparison of performance between an F18-FDG PET normal brain template and a commercial template using the MIMneuro software
Chanisa Chotipanich, Natdanai Hirata, Attapon Jantarato, Peerapon Kiatkittikul, Dheeratama Siripongsatian, Anchisa Kunawudhi, Chetsadaporn Promteangtrong, Nattakoon Tieojaroenkit, Saiphet Vanprom, Nithi Mahanonda
Author Affiliations +
Abstract

Purpose

The aim of this study was to create and validate a normal brain template of F18-fluorodeoxyglucose (F18-FDG) uptake using the MIMneuro software to improve clinical practice.

Approach

One hundred and nine volunteers underwent an F18-FDG positron emission tomography/computed tomography scan. Sixty-three participants with normal Alzheimer’s disease (AD) biomarkers were used to create a template. A group of 23 participants with abnormal AD biomarkers and an additional group of 23 participants with normal AD biomarkers were used to validate the performance of the generated template. The MIMneuro software was used for the analysis and template creation. The performance of our newly created template was compared with that of the MIMneuro software template in the validation groups. Results were confirmed by visual analysis by nuclear medicine physicians.

Results

Our created template provided higher sensitivity, specificity, positive predictive value, and negative predictive value (NPV; 90%, 97.83%, 100%, and 100%, respectively) than did the MIMneuro template when using the positive validation group. Similarly, slightly higher performance was observed for our template than for the MIMneuro template in the negative validation group (the highest specificity and NPV were 100% and 100%, respectively).

Conclusions

Our normal brain template for F18-FDG was shown to be clinically useful because it enabled more accurate discrimination between aging brain and patients with AD. Thus, the template may improve the accuracy of AD diagnoses.

© 2022 Society of Photo-Optical Instrumentation Engineers (SPIE)
Chanisa Chotipanich, Natdanai Hirata, Attapon Jantarato, Peerapon Kiatkittikul, Dheeratama Siripongsatian, Anchisa Kunawudhi, Chetsadaporn Promteangtrong, Nattakoon Tieojaroenkit, Saiphet Vanprom, and Nithi Mahanonda "Comparison of performance between an F18-FDG PET normal brain template and a commercial template using the MIMneuro software," Journal of Medical Imaging 9(6), 064501 (7 November 2022). https://doi.org/10.1117/1.JMI.9.6.064501
Received: 24 March 2022; Accepted: 12 October 2022; Published: 7 November 2022
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Brain

Positron emission tomography

Neuroimaging

Databases

Visualization

Visual analytics

Diagnostics

Back to Top