1 April 2008 Requirements of MEMS membrane mirrors for focus adjustment and aberration correction in endoscopic confocal and optical coherence tomography imaging instruments
Author Affiliations +
Abstract
We explore the use of deformable membrane mirrors for primary focus control and spherical aberration correction in confocal microscopesand optical coherence tomography instruments that are miniaturizedfor in-situ imaging. Membrane requirements to achieve specified system performance are explored for both high and low numerical aperture regimes. Examples of both confocal imaging and focus-tracking optical coherence tomography using varifocus silicon nitride MEMS mirrors are presented. It is envisioned that future instruments with enhanced-stroke adaptive membrane mirrors will provide high fidelity microscopy and optical coherence tomography for diagnostic imaging of intact, living tissue.
©(2008) Society of Photo-Optical Instrumentation Engineers (SPIE)
David L. Dickensheets "Requirements of MEMS membrane mirrors for focus adjustment and aberration correction in endoscopic confocal and optical coherence tomography imaging instruments," Journal of Micro/Nanolithography, MEMS, and MOEMS 7(2), 021008 (1 April 2008). https://doi.org/10.1117/1.2911024
Published: 1 April 2008
Lens.org Logo
CITATIONS
Cited by 21 scholarly publications and 2 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Mirrors

Monochromatic aberrations

Optical coherence tomography

Confocal microscopy

Imaging systems

Tissues

Objectives

Back to Top