28 September 2023 Determination of tissue oxygen saturation by diffuse reflectance spectroscopy
Laura Lucía Sánchez-Ramos, Beatriz Morales-Cruzado, Francisco Gerardo Pérez-Gutiérrez
Author Affiliations +
Abstract

Significance

Tissue oxygenation is a parameter that allows for determining the health status of human beings. In diabetic patients, it is particularly important to evaluate this parameter as an indicator of microcirculatory problems in the extremities.

Aim

We aim to obtain tissue oxygen saturation from diffuse reflectance measurements.

Approach

A computational algorithm to automate the methodology was implemented with the aim of establishing a medical diagnosis technique that is non-invasive and easy to apply and requires a short evaluation time. Tissue oxygen saturation measurements were performed on a group of volunteers to whom a vascular occlusion was applied. It was observed that, by increasing the applied pressure to the arm of each volunteer, the tissue oxygen saturation progressively decreased.

Results

The results indicate that the developed technique is an effective method for monitoring changes in blood hemodynamics in patients with some type of pathology in which tissue oxygenation is compromised. In addition, the expected behavior of tissue oxygen saturation during a vascular occlusion was obtained.

Conclusions

A methodology to obtain tissue oxygen saturation from diffuse reflectance measurements was successfully developed. It meets the necessary characteristics to be considered a technique for obtaining StO2 because it can be applied in vivo and non-invasively and does not require a high computational cost; thus it is fast and capable of providing an objective and quantifiable evaluation.

1.

Introduction

Diabetic foot syndrome is one of the main complications that frequently affect the quality of life of people with diabetes mellitus. This syndrome includes a set of signs and symptoms linked to various vascular, neurological, and infectious disorders. Among the detection methods for an early diagnosis of diabetic foot, which have the objective of taking preventive actions, is the podiatric evaluation that consists of a subjective test known as the Semmes Weinstein monofilament, which evaluates diabetic peripheral neuropathy.1 There is also the ankle brachial index test, which is used to compare blood flow in the arteries of the arm and ankle and represents an indicator of vascular disrepair.2 Another method used to evaluate vascular deterioration is Doppler ultrasound, which uses a transducer to emit sound waves that bounce off blood vessels to create images of the state of those vessels.3 Although these methods can provide important information about the condition of patients with diabetic foot, they generally require a long test time and, in some cases, these do not provide a quantitative result of the analyzed condition.

Optical technologies have emerged as promising tools for medical diagnosis with several advantages such as non-ionizing radiation, low cost, and high molecular and biochemical specificity.4 One of the main applications of optical technologies is oximetry, which is used to measure the oxygenation level of the blood through optical measurements of the transmission/reflection of light through the blood, which is based on the different absorption spectra of oxygenated (HbO2) and deoxygenated (Hb) hemoglobin.

Spectroscopy is one of the techniques used for the evaluation of blood microcirculation because it allows for the analysis of the optical properties of the tissue. Because biological molecules have unique absorption spectra as a function of the wavelength of light, it is possible to detect a precise concentration of these molecules by spectroscopy. Research applying various optical spectroscopy techniques, mainly in the advanced stages of diabetic foot, has been conducted this includes hyperspectral imaging,5 diffuse photon density wave,6 frequency-domain diffuse near-infrared spectroscopy,7 non-contact near-infrared imaging system,8 and terahertz reflection.9 Although most studies focus on the implementation of near infrared spectroscopy technology, there are certain disadvantages such as the positioning of the optical probes and the separation between the emitter and receiver as this separation distance influences the penetration depth of the light beam. Therefore, a technique that provides an objective and quantifiable evaluation is required.

Because blood microcirculation is affected in patients with diabetic foot syndrome, tissue oxygenation is one of the most relevant parameters during the disease process. Because foot ulcers occur when the lack of tissue oxygen prevents cell regeneration, non-invasive quantitative measurements of hemoglobin concentration and tissue oxygen saturation can provide information for diagnostic and therapeutic decision making. Currently, there is no optical technique that is easy to use, reliable, and economical to monitor tissue oxygenation, so this work proposes to cover these needs. The technique is based on diffuse reflectance spectroscopy, which is an especially interesting non-invasive optical technique because the spectrum acquired to apply this technique presents information about the chromophores (absorbent elements) of the tissue and its morphology. Because blood is the dominant light absorber in the visible wavelength range, both local blood content and tissue oxygen saturation can be extracted from the reflected light signal using an integrating sphere. The results of the work presented here demonstrate that it is feasible to implement a portable device to monitor tissue oxygen saturation in clinical practice. Such a device works with light emitting diode (LED) lighting and detection with photodiodes.

2.

Methodology

2.1.

Oxygen Saturation

Due to the critical nature of tissue oxygen consumption in the body, oxygen saturation, the amount of hemoglobin bound to oxygen compared with the amount of unbound hemoglobin, is an essential element of patient care; thus it is currently considered to be the fifth vital sign.10

Blood oxygenation can be evaluated by monitoring oxygenated hemoglobin (HbO2) and deoxygenated or reduced hemoglobin (Hb) concentration parameters, which are also used to calculate the SO2 oxygen saturation parameter as

Eq. (1)

SO2=[HbO2][HbO2]+[Hb]x100%,
where oxygen saturation SO2 represents the concentration of oxygenated hemoglobin in relation to total hemoglobin. SO2 takes different names depending on where it is measured. For example, SaO2 for artery, SvO2 for veins, and StO2 for tissue. Thus, StO2 is the ratio between oxygenated hemoglobin and total hemoglobin in all tissue blood: arterioles, capillaries, and venules. It is directly related to tissue blood flow and inversely related to its metabolic requirements.

Kyriacou et al.11 present the molar extinction coefficients or specific absorption constants of HbO2 and Hb as a function of wavelength, in the visible and near-infrared wavelength regions, where a large difference between the extinction coefficients of HbO2 and Hb is clear for some wavelengths. Another important aspect is that HbO2 and Hb have a different absorption throughout the spectrum, except at 805 nm, which is called the isosbestic point. For shorter wavelengths, the absorption is mainly due to Hb, whereas for longer wavelengths, most of the absorption is due to HbO2.

To obtain the oxygen saturation value using Eq. (1), it is necessary to select at least two wavelengths considering three main criteria: first, the selected wavelengths (λ1 and λ2) must be sensitive to each of the main blood absorbers, HbO2 and Hb (which are generally opposite the isosbestic point). Second, the absorption difference between HbO2 and Hb at the chosen wavelengths should be the largest. Finally, the absorption spectra in the region of the chosen wavelengths should be relatively flat, indicating their proximity in terms of absorption magnitude.11

2.2.

Propagation of Optical Radiation in Human Skin

The interval of the electromagnetic spectrum in which the tissues have a significant level of light penetration to provide information on the state of the tissues, but without causing a harmful effect on health, is known as the therapeutic window and is located between the wavelengths of 600 and 1300 nm.12

The interactions of electromagnetic radiation with the skin are governed by its physiological, biochemical, and morphological characteristics.13 They are characterized by a strong optical scattering due to the high degree of inhomogeneity caused by the organelles that the cells contain, which have different sizes and compositions; likewise they present a different refractive index to that of the substance in which the fibers are immersed and connective tissue cells.14 Therefore, the skin can be considered to be an inhomogeneous turbid medium.

When electromagnetic radiation interacts with the skin, a part of the radiation is reflected due to the difference between the refractive indices of the skin and the air, and the remaining radiation that crosses the border can be absorbed internally or scattered multiple times by the middle. Radiation that has undergone multiple scattering events can be diffusely transmitted or reflected.15 To describe the propagation of electromagnetic radiation in the skin, the absorption coefficient (μa), scattering coefficient (μs), anisotropy factor (g), and refractive index (n) are used as the most important optical properties.16

Because human skin is a tissue made up of different layers that contain a diversity of non-homogeneous structures, the study of the interaction of electromagnetic radiation with this medium becomes difficult due to the diversity in the size, shape, density, and refractive index of these components. Therefore, simplifications are generally made for their study, and the structures are considered to be homogeneously distributed spherical particles.17

2.3.

Radiative Transport Theory

The radiative transport theory considers that the only interaction of light with the medium is of an elastic type, changing the energy distribution of the incident light as it passes through such a medium.1821 This theory does not include the effects of coherence, polarization, interference, and diffraction. Its fundamental equation is the radiative transfer equation (RTE), which is an integral–differential equation. The radiance or specific intensity is the unknown quantity in this equation and includes parameters such as the scattering and absorption coefficients and the phase function of the medium.

To solve the RTE, the main problem that arises is the evaluation of the diffuse radiance because the scattered photons do not follow a determined trajectory. For this reason, appropriate statistical approximations and approaches must be chosen, that is, whether absorption or scattering is the dominant attenuation process.

The methods used are generally the Kubelka–Munk theory, the diffusion approximation, the Monte Carlo method, or the adding–doubling method. Each method is based on certain approximations that facilitate the RTE solution. In general, the complexity of any of the approaches is closely related to its precision, as well as to the computation time required to give a solution.

The inverse adding–doubling (IAD) method provides a tool for the fast and accurate solution of the inverse scattering problem. Adding–doubling is based on the general method for the solution of the RTE for flat and parallel layers, suggested by Van de Hulst and introduced to the optics of biological tissues by Prahl et al.22

An important advantage of the IAD method when applied to biological tissue optics is the possibility of rapidly obtaining iterative solutions. Furthermore, it is flexible enough to consider scattering anisotropy and internal reflection from sample boundaries.

2.4.

Experimental Set-Up

To obtain tissue oxygen saturation, it is necessary to obtain the concentrations corresponding to oxygenated (HbO2) and deoxygenated (Hb) hemoglobin. Those concentrations can be determined from the absorption coefficients of the tissue to be analyzed. The absorption coefficients are recovered by IAD from diffuse reflectance measurements made with an integrating sphere experimental set-up (see Fig. 1). The experimental set-up was used to measure the reflected light by the sample; then it is normalized and used as input parameter together with the scattering coefficient and the anisotropy factor.23

Fig. 1

Experimental set-up for diffuse reflectance measurement. Inset: integrating sphere used to measure diffuse reflectance of the volunteer hand.

JBO_28_9_095002_f001.png

The experimental set-up consists of a tungsten lamp (HL-2000; Ocean Optics) that emits radiation in the visible spectrum (even though the spectrum is specified from 360 to 1700 nm) coupled to an optical fiber that carries light from the lamp to the input port of an integrating sphere (819C-SF-6; Newport) passing through a collimating lens (fiber optics do not penetrate the integrating sphere). The collimated light reaches the sample, which is placed in the opposite port to the entrance of the light source and is reflected in a specular and diffusive way. An optical fiber connected to the top port of the integrating sphere carries the diffusively reflected light from the sample to a spectrometer (USB4000; Ocean Optics, with a detection range between 200 and 1100 nm). A continuous wave collimating laser (BWF-975-450E) with a central wavelength of 975 nm and a maximum output power of 450 mW was used as the light source. For the experiments, only 1.0 mW was used in an area of 2.27  mm2, which is harmless for human skin.

The reflectance measurement was performed by placing the palm of the hand on the outlet port of the integrating sphere, as shown in Fig. 1. The light reflected by the tissue is collected and guided to the spectrometer by means of an optical fiber, and the spectrometer separates the reflected light at different wavelengths and provides a spectrum of the intensity.

The reflection measurements are normalized with respect to the incident light. Applying Eq. (2), a value of the total diffuse reflectance is obtained as

Eq. (2)

MR=rstd·RRR0R1R0+RFresnel,
where rstd represents the reflection factor of the integrating sphere and the reference standard, RR is the reflection measured from the palm of the volunteer’s hand, R0 is equivalent to the minimum reflection that can be measured and represents the electronic noise that is measured leaving the output port empty, and R1 is the maximum reflection that can be measured and corresponds to the spectrum of the lamp that is measured by placing the reflection standard on the output port. Measurements R0, R1, and RR are the result of averaging several spectrophotometer readings with an integration time of 100 ms and a boxcar width of 50. The combination of both factors, integration time and boxcar width, with the amount of light emitted by the sources provides a high signal to noise ratio. The term RFresnel refers to the Fresnel or specular reflection produced by the difference between the refractive indices of air and skin. Figure 2 shows the schematic representations of the measurements of R0, R1, and RR.

Fig. 2

Schematic representations of the measurement of R0, R1, and RR for calculation the of MR.

JBO_28_9_095002_f002.png

The experimental tests were performed on healthy volunteers; for this reason, it was necessary to induce ischemia to simulate the reduction in blood flow that occurs in diabetic patients with peripheral vascular disease. A vascular occlusion test was performed using a sphygmomanometer; the blood pressure cuff was placed around the upper left arm and inflated to stop the flow of blood in the arteries/veins up to a certain pressure value indicated on the manometer. The vascular occlusion test has been used in various studies such as those carried out by Wang et al.,24 who applied the occlusion test to healthy volunteers to evaluate their hybrid diffuse optical device in stroke diagnosis. Donati et al.25 used the occlusion test in 27 healthy volunteers to detect alterations in tissue oxygenation and microvascular reactivity with the aim of predicting adverse outcome in critically ill patients. Bezemer et al.26 applied the test to eight healthy volunteers to assess potential metabolic and microcirculatory alterations that patients may suffer during sepsis or shock.

The study followed the tenets of the Declaration of Helsinki and the Mexican General Health Law on Research for Health, and the corresponding protocol was approved by the Ethics and Research Committee CONBIOETICA of the UASLP (registration number CEI-FE-046-023). The volunteers were informed in detail about the experimental procedure, and they provided written consent for their data to be used for research purposes. The group of participants consisted of 27 healthy volunteers, 13 men and 14 women, between 18 and 39 years old. For the experimental tests, volunteers were asked to sit comfortably in a chair and rest for 3 min. During this time, blood pressure, heart rate, and pulsatile oxygen saturation were measured. In addition, a diffuse reflection spectrum was taken from each volunteer to calculate their absorption spectrum (see Supplementary Material). The absorption coefficient at 632 nm was correlated with melanin content, according to a prior study27 and classified into skin type on the Fitzpatrick scale,28 with the volunteers being between skin types III and V. Subsequently, a sphygmomanometer around the left arm was placed to induce vascular occlusions, and the palm of their left hand was placed on the port of the integrating sphere, as shown in Fig. 3.

Fig. 3

Setting of a sphygmomanometer for measurements of StO2.

JBO_28_9_095002_f003.png

Four continuous measurements were made at 0, 60, 80, and 100 mmHg of the 27 adult volunteers. Each occlusion took 1  min. At the end of the 100 mmHg measurement, the pressure in the cuff was released, and measurements were taken in the recovery period 1 and 2 min later. Volunteer characteristics such as age, blood pressure, heart rate, pulsatile oxygen saturation, and skin type were registered. The information of each volunteer was tagged using a consecutive number followed by their gender in parentheses.

2.5.

Algorithm for Obtaining Oxygen Saturation

The tissue absorption coefficient μa can be considered to be the sum of n products of the molar extinction coefficients of the significant chromophores εi and their respective concentrations Ci, where (λ) indicates a wavelength dependence, as calculated in Eq. (3):

Eq. (3)

μa(λ)=i=1nεi(λ)Ci.

From Eq. (2), concentrations of oxygenated hemoglobin HbO2 and deoxygenated or reduced hemoglobin Hb were obtained from the tissue absorption coefficients μa and molar extinction coefficients ε corresponding to each hemoglobin in two wavelengths λ1 and λ2, as indicated in Eqs. (4) and (5):

Eq. (4)

[HbO2]=μaλ2εHbλ1μaλ1εHbλ2εHbλ1εHbO2λ2εHbλ2εHbO2λ1,

Eq. (5)

[Hb]=μaλ1εHbO2λ2μaλ2εHbO2λ1εHbλ1εHbO2λ2εHbλ2εHbO2λ1.
where HbO2 and Hb have characteristic absorption spectra, reported on the Oregon Medical Laser Center website.29 These spectra correspond to the molar extinction coefficients of oxygenated εHbO2 and deoxygenated εHb hemoglobin as a function of wavelength for the visible and near infrared regions, where the molar extinction coefficient is defined as an absorption constant per molar unit.

Due to the difference between εHbO2 and εHb identified in the visible and near infrared region, these coefficients can be used for the evaluation of blood oxygenation.30 The chosen wavelengths to obtain the StO2 in this work were 975 nm in the infrared region, as well as a wavelength in the visible region positioned at a similar distance from the isosbestic point as the infrared point and showing significant changes in the StO2.

The absorption coefficients μaλ1 and μaλ2 in Eqs. (3) and (4) were obtained from the in vivo measurements of diffuse reflectance with the integrating sphere experimental set-up and following the procedure described by Quistian-Vázquez et al.23 to obtain the absorption coefficient μa. The input parameters were the measurement of the diffuse reflectance MR, the scattering coefficient μs, and the anisotropy factor g of the skin; these depend on the biochemical composition of the tissue and its cellular structure, as well as on the wavelength.

The values of μs as a function of wavelength were obtained from Jacques’ review.31 Equation (6) characterizes scattering in seven tissue groups: skin, brain, breast, bone, soft tissues, fibrous tissues, and fatty tissues. The values of a and b for skin are a=46  cm1 and b=1.421, respectively,

Eq. (6)

μs(λ)=a(λ500  nm)b.

To define g(λ) as a function of the wavelength, a sixth order polynomial was determined; it represents a polynomial fit of the values obtained by Ma et al.32 in a study carried out to determine the optical properties of the skin and that are presented in the Jacques review.31 Equation (7) shows the function obtained for g(λ):

Eq. (7)

g(λ)=5.603+3.61×102·λ8.17×105·λ2+9.51×108·λ35.92×1011·λ4+1.83×1014·λ52.11×1018·λ6.

The scattering coefficient μs(λ) as a function of the wavelength was obtained from the relation μs=μs(1g).

3.

Computer Program: Obtaining Tissue Oxygen Saturation

The algorithm “obtaining tissue oxygen saturation (OTOS)” was developed, taking as a reference the program developed by Quistian-Vázquez et al.,23 which uses a series of routines ranging from optical spectrometer reading until the retrieval of the absorption or scattering optical properties applying IAD. The Quistian-Vázquez program implemented in vivo measurements to monitor the absorption coefficient in real time. In a study carried out to evaluate the effect of temperature, results demonstrated a temperature dependence on the optical properties of human skin, which is important in the prescription of light doses based on optical properties. These results indicate the feasibility of using the integrating sphere in measurements of optical properties in vivo to monitor the state or health condition of people. On the other hand, Rehman et al.33 stated that the integrating sphere system together with the IAD method can differentiate between normal and diseased skin tissue based on the contrast in scattering and absorption properties.

To obtain the absorption coefficient μa, the measurement of diffuse reflectance is required as input parameter MR and the optical properties of human skin as scattering coefficient μs and anisotropy factor g. The values of μs and g as a function of wavelength were obtained from Eqs. (6) and (7), respectively. Subsequently, the wavelength range from 600 to 700 nm was defined to retrieve the absorption coefficient in the visible region of the spectrum and to determine a linear fit in the range of 630 to 700 nm. This slope is related to the melanin content present in the skin, as demonstrated by Kollias and Baqer34 and Stamatas et al.35 Routines for reading the molar extinction coefficients of oxygenated and deoxygenated hemoglobin and to calculate tissue oxygen saturation were also included in the computer program. The values of εHbO2 and εHb considered in this work are those reported by Prahl.29

The graphical interface, as shown in Fig. 4, allows the user to define the input values for the spectrometer configuration (parts 1–3), as well as the properties of the integrating sphere (part 4). It also allows for viewing the values of the absorption/extinction coefficients and tissue oxygen saturation, both in the visible region and in the infrared point, as observed in parts 5–9.

Fig. 4

Parts of the graphical interface corresponding to the different stages of the algorithm for OTOS.

JBO_28_9_095002_f004.png

The implemented algorithm to obtain tissue oxygen saturation allows for automating the measurements because it optimizes the processing time between each measurement. The implementation of the graphical interface facilitates the application of the algorithm by anyone because it does not require specific technical training. In this way, the developed algorithm provides a good tool for use in clinical practice, such as is routine in appointments for the control and follow-up of diabetic patients, for which medical diagnostic techniques that take up less evaluation time, are not invasive, and are easy to apply are required.

4.

Results and Discussion

4.1.

Measurement of Absorption Coefficient Spectra In Vivo

Absorption spectra from diffuse reflectance measurements were obtained using the experimental set-up shown in Fig. 1 and applying the IAD technique. The graphs of Fig. 5 show the absorption coefficients spectra of 2 volunteers for each condition of applied pressure (0, 60, 80, and 100 mmHg) in which it is observed that the absorption coefficient increases with increasing pressure in the sphygmomanometer because of the increase in the concentration of deoxygenated hemoglobin (Hb). Figure 5(a) shows that, by increasing the applied pressure, the characteristic double peak pattern of oxygenated hemoglobin (HbO2) is replaced by the characteristic single peak pattern of deoxygenated hemoglobin (Hb). This behavior is better appreciated in Fig. 5(b) corresponding to a volunteer with a lower systolic pressure. In Fig. 5, the spectra also depend on the skin type because darker skin presents greater absorption.36

Fig. 5

(a) and (b) Absorption coefficients of two volunteers for each condition of applied pressure.

JBO_28_9_095002_f005.png

There are several factors that can affect the measurements, which can be classified as anatomical or physiological; however, a statistical study of experimental data by principal component analysis (PCA) of the main skin chromophores (considering the absorption coefficients of melanin, deoxygenated hemoglobin, oxygenated hemoglobin, water, and fat), revealed that the chromophores that have the greatest effect on the total absorption coefficient are oxygenated and deoxygenated hemoglobin and melanin. PCA details can be found in the Supplementary Material.

4.2.

In Vivo Tissue Oxygen Saturation Measurement with OTOS Algorithm

The StO2 measurements were carried out with the developed computational algorithm. The skin type was classified according to three groups: I–fair skin, II–intermediate skin, and III–brown skin.

Subsequently, diffuse reflectance measurements and absorption coefficients were obtained for each pressure condition (0, 60, 80, and 100 mmHg). When obtaining the measurement of 0 mmHg (basal or resting condition), the slope of the absorption spectrum between 630 and 700 nm was calculated because that slope is directly related to the melanin content in the skin.35 The slope obtained for each volunteer was used to discriminate the contribution of epidermal melanin to the absorption spectrum. From the value of the slope, the first wavelength necessary to perform the calculation of StO2, corresponding to the visible region of the electromagnetic spectrum, was chosen. The second wavelength corresponding to the near infrared region was 975 nm due to the availability of the infrared (IR) laser. The three groups specified for the skin type were related to the obtained slope value because the skin type is directly associated with the melanin content in the skin. Table 1 shows the classification group, as well as the interval of the corresponding slope value and selected wavelength for each group.

Table 1

Wavelength determination for StO2 calculation (considering the absolute value of the slope).

GroupSkin typeAbsolute value of slope 630 to 700 nm (mm−1/nm)Wavelength (nm).
IFairSlope<5×104622
IIIntermediate5×104<slope<8×104624
IIIBrown8×104<slope626

To assign a wavelength for each group of skin type, the behavior indicated by Ferguson-Pell and Hagisawa37 was considered. It was observed that there was a change in wavelength in the calculation of the oxygenation index due to the influence of melanin. The obtained values of StO2 for each pressure condition during the occlusion test are shown in Figs. 6Fig. 78, where it can be seen that the value of StO2 tends to decrease as the applied pressure increases. Measurements were categorized in groups by gender, age, and corresponding systolic blood pressure measured before test. Figure 6(a) shows the results belonging to the group of women, and Fig. 6(b) corresponds to the men. Regarding gender, no correlation was found, so it can be concluded that it does not influence the measurement of tissue oxygen saturation.

Fig. 6

Graphs of StO2 values obtained for each pressure condition of occlusion test classified by gender: (a) female and (b) male.

JBO_28_9_095002_f006.png

Fig. 7

Graphs of StO2 for each pressure condition of occlusion test classified by age: (a) 18 to 20 years old, (b) 20 to 30 years old, and (c) 30 to 40 years old.

JBO_28_9_095002_f007.png

Fig. 8

Graphs of StO2 values obtained during the occlusion test: systolic blood pressure (a) 90 to 104 mmHg, (b) 105 to 114 mmHg, and (c) 115 to 130 mmHg.

JBO_28_9_095002_f008.png

The classification by age is shown in Fig. 7. The results for the 10 to 20 years old group is observed in Fig. 7(a), for the 20 to 30 years old group in Fig. 7(b), and the 30 to 40 years old group is in Fig. 7(c). No special correlation between volunteers of a similar age was observed. Age influences some physiological characteristics such as hydration and the quantity/structure of collagen fibers present in the skin. However, the volunteers’ ages ranged between 18 and 39 years, which does not represent a considerable difference.

Figure 8(a) shows the results for the group that presented a systolic blood pressure between 90 and 104 mmHg, Fig. 8(b) corresponds to the group with a blood pressure of 105 to 114 mmHg, and Fig. 8(c) shows the group with a blood pressure of 115 to 130 mmHg.

Classifications by gender and age do not present a special pattern or behavior. However, in the classification by group of systolic blood pressure, it is observed that the group with the lowest blood pressure (90 to 104 mmHg) presents a pronounced drop in StO2, unlike the group with the highest blood pressure (115 to 130 mmHg). This is an expected behavior because the volunteers with lower systolic blood pressure practically reach a situation of zero blood flow in the highest applied pressure (100 mmHg); therefore they present a greater affectation. It is also expected that, in the absorption spectra of this group of volunteers, the characteristic double peak pattern of oxygenated hemoglobin gets lost faster, as is the case of volunteer V21(M) in Fig. 8(a) which corresponds to volunteer V7(M) in Fig. 5(b).

Figure 9 shows the graphs of StO2 versus time in minutes, where 1 measurement was taken each 1 min. These graphs also show the measurements corresponding to the StO2 recovery stage, that is, after evaluating 100 mmHg, the pressure applied by the sphygmomanometer cuff was immediately released and 2 consecutive measurements were taken.

Fig. 9

Graphs of StO2 values versus test time: volunteer’s initial systolic blood pressure (a) 90 to 104 mmHg, (b) 105 to 114 mmHg, and (c) 115 to 130 mmHg.

JBO_28_9_095002_f009.png

In Fig. 9, it can be seen that the StO2 tends to go up again in the recovery stage, and a heterogeneous behavior can be observed between the recovery stage of the different volunteers of each group. Based on observations made in various studies, this recovery stage is considered to be a reflection of the endothelial function and has been correlated with perfusion pressure; therefore, it can be derived from the interaction between perfusion pressure and endothelial integrity.38 However, the obtained results indicate a much lower ischemic challenge than that study because the maximum applied pressure (100 mmHg) was not as high as the initial systolic blood pressure and therefore does not present a good degree of comparison. Despite this, in the volunteers V12(H) in Fig. 9(b) and V25(M) in Fig. 9(c) it was observed that, in the first measurement of the recovery stage, the StO2 temporarily went up above the initial values. This behavior may indicate the condition known as reactive hyperemia, mentioned in the literature, which represents post-ischemic vasodilation and capillary recruitment.38,39 Although multiple comparisons were made to identify differences between subgroups of volunteers, no statistically significant differences between gender, age, or skin color were found.

The validation of the results obtained in comparison with values from the literature is difficult because there is not a defined standard and previous studies report different results regarding the measurement zones, the separation between the optical probes, and the algorithms used. However, our results showed agreement with those reported by Tseng et al.,40 in which they performed diffuse reflectance measurements with visible and near-infrared light and determined two regions to obtain the StO2. The region between 500 and 600 nm provide saturation values between 60% and 70% because the depth of penetration is lower and the measurements correspond to the blood vessels of the upper dermis, whereas the region from 600 to 1000 nm provides values of saturation from 97% to 99% by covering a greater depth of measurement in the region of the lower dermis.

The proposed method gives saturation values between 97% and 99% for healthy people under normal conditions; therefore it satisfies the necessary characteristics to be considered a technique for obtaining StO2 because it can be applied in vivo and non-invasively, requires less evaluation time, and provides an objective and quantifiable evaluation.

As future work, the implementation of the portable device designed to apply the technique in clinical practice will be considered; it will work with LED illumination and photodiodes for detection. In addition, this device has the advantage of changing the size of the port and thus being able to measure smaller areas, allowing for the application of the technique to other possible conditions.

5.

Conclusions

A technique for measuring tissue oxygen saturation (StO2) was developed with the aim of monitoring microcirculatory changes in patients at risk of suffering from diabetic foot. The developed technique used an integrating sphere as a measurement device, which reduced the problems of positioning and separation of optical probes; it constituted a portable and flexible measurement method in various applications. The developed procedure measured the diffuse reflectance spectra of the human skin and used the IAD algorithm to obtain the absorption coefficients to obtain the StO2 values. The computer program OTOS was developed to automate the implemented methodology and thus facilitate its use in clinical applications.

The developed OTOS program was applied with a vascular occlusion test to assess changes in StO2 in a group of 27 volunteers. The occlusion test consisted of placing a sphygmomanometer on the volunteers’ left arms and progressively increasing the pressure to 0, 60, 80, and 100 mmHg while the absorption coefficients in the palm of the left hand were measured. The results showed the expected behavior, according to the information reported in the literature. By increasing the applied pressure, the StO2 also decreased by an average of 95.69%, 94.78%, 92.45%, and 91.46%, respectively. It was observed that the volunteers with lower systolic pressures had a greater drop in StO2 because the applied pressure of 100 mmHg was above the systolic blood pressure in some of the volunteers. In the measurement of the absorption spectra, it was observed that, by increasing the applied pressure, the characteristic double peak pattern of oxygenated hemoglobin between 500 and 600 nm was replaced by the single peak pattern that characterizes deoxygenated hemoglobin; this behavior indicated that the blood began to deoxygenate.

A statistical study of experimental data by PCA was developed; it was observed that oxygenated and deoxygenated hemoglobin and melanin are the most significant chromophores in the absorption spectrum in the wavelength range under study. However, the developed algorithm is capable of eliminating the effect of melanin by correlating the slope of the absorption spectrum in the interval from 630 to 700 nm with the melanin content; therefore the value of StO2 did not depend on the skin type of the person on whom the measurement was made.

From the obtained results, we can conclude that the proposed method meets the necessary characteristics to be considered a technique for obtaining StO2 because it can be applied in vivo and non-invasively and does not require a high computational cost; thus it is fast and capable of providing an objective and quantifiable evaluation. For patients at risk of developing diabetic foot, the proposed method can help quantify the decrease in StO2 as an indication of the reduction in limb blood flow during follow-up medical visits.4148

Disclosures

The authors declare no conflicts of interest with the research presented in this paper.

Code, Data, and Materials Availability

Code and data will be made available on request.

Acknowledgments

The authors acknowledge funding from Mexico’s National Council of Humanities, Science, and Technology (CONACYT; grant nos. A1-S-9887 and CF-2019-102986).

References

1. 

L. S. Tan, “The clinical use of the 10g monofilament and its limitations: a review,” Diabetes Res. Clin. Pract., 90 (1), 1 –7 https://doi.org/10.1016/j.diabres.2010.06.021 DRCPE9 0168-8227 (2010). Google Scholar

2. 

J. C. Sánchez-Ruiz et al., “Utilidad del índice tobillo-brazo en atención primaria,” Med. Fam. – SEMERGEN, 31 (11), 533 –535 https://doi.org/10.1016/S1138-3593(05)72985-3 (2005). Google Scholar

3. 

O. B. Suade et al., “Arterial evaluation of systemic sclerosis patients with Doppler ultrasound: what did we find?,” Int. J. Rheumat. Dis., 24 (7), 954 –962 https://doi.org/10.1111/1756-185X.14158 (2021). Google Scholar

4. 

R. G. Frykberg et al., “Diabetic foot disorders: a clinical practice guideline,” J. Foot Ankle Surg., 45 (5), S1 –S66 https://doi.org/10.1016/S1067-2516(07)60001-5 (2006). Google Scholar

5. 

R. L. Greenman et al., “Early changes in the skin microcirculation and muscle metabolism of the diabetic foot,” Lancet, 366 (9498), 1711 –1717 https://doi.org/10.1016/S0140-6736(05)67696-9 LANCAO 0140-6736 (2005). Google Scholar

6. 

E. S. Papazoglou et al., “Noninvasive assessment of diabetic foot ulcers with diffuse photon density wave methodology: pilot human study,” J. Biomed. Opt., 14 (6), 064032 https://doi.org/10.1117/1.3275467 JBOPFO 1083-3668 (2009). Google Scholar

7. 

J. A. Samuels et al., “Diffuse near infrared spectroscopy assessment of diabetic foot ulcers: a human study. Biomedical optics and 3D imaging,” in OSA Tech. Digest, JM3A.34 (2012). Google Scholar

8. 

A. Godavarty et al., “Non-contact optical imaging of healing and non-healing diabetic foot ulcers,” Proc. SPIE, 9318 931802 https://doi.org/10.1117/12.2080716 PSISDG 0277-786X (2015). Google Scholar

9. 

G. G. Hernandez-Cardoso et al., “Terahertz imaging for early screening of diabetic foot syndrome: a proof of concept,” Sci. Rep., 7 42124 https://doi.org/10.1038/srep42124 (2017). Google Scholar

10. 

B. B Hafen and S. Sharma, “Oxygen saturation,” https://www.ncbi.nlm.nih.gov/books/NBK525974/ (2022). Google Scholar

11. 

P. Kyriacou, K. Budidha and T. Y. Abay, “Optical techniques for blood and tissue oxygenation,” Encyclopedia Biomed. Eng., 3 461 –472 https://doi.org/10.1016/B978-0-12-801238-3.10886-4 (2019). Google Scholar

12. 

G. Lu and B. Fei, “Medical hyperspectral imaging: a review,” J. Biomed. Opt., 19 (1), 010901 https://doi.org/10.1117/1.JBO.19.1.010901 JBOPFO 1083-3668 (2014). Google Scholar

13. 

Y. Ti and W. C. Lin, “Effects of probe contact pressure on in vivo optical spectroscopy,” Opt. Express, 16 4250 –4262 https://doi.org/10.1364/OE.16.004250 OPEXFF 1094-4087 (2008). Google Scholar

14. 

V. Tuchin, Tissue Optics: Light Scattering Methods and Instruments for Medical Diagnosis, 3rd ed.SPIE Press, Bellingham, Washington, United States (2015). Google Scholar

15. 

A. Karsten et al., “Diffuse reflectance spectroscopy as a tool to measure the absorption coefficient in skin: system calibration,” Lasers Med. Sci., 28 (2), 437 –444 https://doi.org/10.1007/s10103-012-1079-2 LMSCEZ 1435-604X (2013). Google Scholar

16. 

A. J. Welch and M. J. C. Gemert, Optical-Thermal Response of Laser Irradiated Tissue, Springer Science( (1995). Google Scholar

17. 

J. Campoy, R. Alcarria and A. P. González-Marcos, “Estudio del comportamiento óptico de pieles utilizando técnicas de Monte Carlo,” Rev. Int. Metod. Numer., 31 (3), 161 –170 https://doi.org/10.1016/j.rimni.2014.04.004 (2015). Google Scholar

18. 

L. V. Wang and H. Wu, Biomedical Optics: Principles and Imaging, John Wiley & Sons( (2012). Google Scholar

19. 

A. Ishimaru, Wave Propagation and Scattering in Random Media, Academic Press & Oxford University( (1997). Google Scholar

20. 

M. H. Niemz, Laser-Tissue Interactions: Fundamentals and Applications, 4th ed.Springer( (2019). Google Scholar

21. 

D. A. Boas, C. Pitris and N. Ramanujam, Handbook of Biomedical Optics, CRC Press( (2011). Google Scholar

22. 

S. A. Prahl, M. J. C. Gemert and A. Welch, “Determining the optical properties of turbid media by using the adding-doubling method,” Appl. Opt., 32 (4), 559 –568 https://doi.org/10.1364/AO.32.000559 APOPAI 0003-6935 (1993). Google Scholar

23. 

B. Quistián-Vázquez et al., “Retrieval of Absorption or Scattering Coefficient Spectrum (RASCS) program: a tool to monitor optical properties in real time,” Lasers Surg. Med., 52 (6), 552 –559 https://doi.org/10.1002/lsm.23164 LSMEDI 0196-8092 (2020). Google Scholar

24. 

D. Wang et al., “Simultaneous in vivo measurements of the total hemoglobin, oxygen saturation, and tissue blood flow via hybrid near-infrared diffuse optical techniques,” AIP Adv., 9 065306 https://doi.org/10.1063/1.5095699 AAIDBI 2158-3226 (2019). Google Scholar

25. 

A. Donati et al., “Near-infrared spectroscopy for assessing tissue oxygenation and microvascular reactivity in critically ill patients: a prospective observational study,” Crit. Care, 20 (1), 311 https://doi.org/10.1186/s13054-016-1500-5 (2016). Google Scholar

26. 

R. Bezemer et al., “Assessment of tissue oxygen saturation during a vascular occlusion test using near-infrared spectroscopy: the role of probe spacing and measurement site studied in healthy volunteers,” Crit. Care, 13 (5), 1 –7 https://doi.org/10.1186/cc8002 (2009). Google Scholar

27. 

A. E. Karsten and J. E. Smit, “Modeling and verification of melanin concentration on human skin type,” Photochem. Photobiol., 88 469 –474 https://doi.org/10.1111/j.1751-1097.2011.01044.x PHCBAP 0031-8655 (2012). Google Scholar

28. 

T. B. Fitzpatrick, “The validity and practicality of sun-reactive skin types I through VI,” Arch. Dermatol., 124 869 –871 https://doi.org/10.1001/archderm.1988.01670060015008 (1988). Google Scholar

29. 

S. Prahl, “Optical absorption of hemoglobin,” https://omlc.org/spectra/hemoglobin/ (). Google Scholar

30. 

M. Nitzan, I. Nitzan and Y. Arieli, “The various oximetric techniques used for the evaluation of blood oxygenation,” Sensors, 20 (17), 4844 https://doi.org/10.3390/s20174844 SNSRES 0746-9462 (2020). Google Scholar

31. 

S. L. Jacques, “Corrigendum: optical properties of biological tissues: a review,” Phys. Med. Biol., 58 (14), 5007 https://doi.org/10.1088/0031-9155/58/14/5007 PHMBA7 0031-9155 (2013). Google Scholar

32. 

X. Ma et al., “Bulk optical parameters of porcine skin dermis at eight wavelengths from 325 to 1557 nm,” Opt. Lett., 30 (4), 412 –414 https://doi.org/10.1364/OL.30.000412 OPLEDP 0146-9592 (2005). Google Scholar

33. 

A. U. I Rehman, I. Ahmad and S. A. Qureshi, “Biomedical applications of integrating sphere: a review,” Photodiagn. Photodyn. Ther., 31 101712 https://doi.org/10.1016/j.pdpdt.2020.101712 (2020). Google Scholar

34. 

N. Kollias and A. Baqer, “Spectroscopic characteristics of human melanin in vivo,” J. Invest. Dermatol., 85 (1), 38 –42 https://doi.org/10.1111/1523-1747.ep12275011 JIDEAE 0022-202X (1985). Google Scholar

35. 

G. N. Stamatas et al., “In vivo measurement of skin erythema and pigmentation: new means of implementation of diffuse reflectance spectroscopy with a commercial instrument,” Br. J. Dermatol., 159 (3), 683 –690 https://doi.org/10.1111/j.1365-2133.2008.08642.x BJDEAZ 0007-0963 (2008). Google Scholar

36. 

B. Quistián-Vázquez et al., “Study of the effect of temperature on the optical properties of Latin skins,” Proc. SPIE, 10062 1006210 https://doi.org/10.1117/12.2252945 PSISDG 0277-786X (2017). Google Scholar

37. 

M. Ferguson-Pell and S. Hagisawa, “An empirical technique to compensate for melanin when monitoring skin microcirculation using reflectance spectrophotometry,” Med. Eng. Phys., 17 (2), 104 –110 https://doi.org/10.1016/1350-4533(95)91880-P MEPHEO 1350-4533 (1995). Google Scholar

38. 

J. Mesquida, G. Gruartmoner and C. Espinal, “Skeletal muscle oxygen saturation (StO2) measured by near-infrared spectroscopy in the critically Ill patients,” BioMed. Res. Int., 2013 502194 https://doi.org/10.1155/2013/502194 (2013). Google Scholar

39. 

G. Gruartmoner, J. Mesquida and F. Baigorri, “Saturación tisular de oxígeno en el paciente crítico,” Med. Intensiva, 38 (4), 240 –248 https://doi.org/10.1016/j.medin.2013.07.004 (2014). Google Scholar

40. 

S. Tseng et al., “Chromophore concentrations, absorption and scattering properties of human skin in-vivo,” Opt. Express, 17 (17), 14599 –14617 https://doi.org/10.1364/OE.17.014599 OPEXFF 1094-4087 (2009). Google Scholar

41. 

W. W. Daniel, Biostatistics: A Foundation for Analysis in the Health Sciences, 3rd ed.John Wiley & Sons( (2018). Google Scholar

42. 

T. Abay and P. Kyriacou, “Comparison of NIRS, laser Doppler flowmetry, photoplethysmography, and pulse oximetry during vascular occlusion challenges,” Physiol. Meas., 37 (4), 503 –514 https://doi.org/10.1088/0967-3334/37/4/503 PMEAE3 0967-3334 (2016). Google Scholar

43. 

V. Hartwig et al., “Assessment of microvascular function using near-infrared spectroscopic 2D imaging of whole hand combined with vascular occlusion test,” J. Med. Biol. Eng., 36 87 –95 https://doi.org/10.1007/s40846-016-0114-3 IYSEAK 0021-3292 (2016). Google Scholar

44. 

R. P. E. Boezeman et al., “In vivo measurements of regional hemoglobin oxygen saturation values and limb-to-arm ratios of near-infrared spectroscopy for tissue oxygenation monitoring of lower extremities in healthy subjects,” Med. Devices: Evidence Res., 8 31 –36 https://doi.org/10.2147/MDER.S73103 (2015). Google Scholar

45. 

J. Fellahi et al., “Lower limb peripheral NIRS parameters during a vascular occlusion test: an experimental study in healthy volunteers,” Ann. Fr. Anesth. Réanim., 33 e9 –e14 https://doi.org/10.1016/j.annfar.2013.11.014 (2014). Google Scholar

46. 

J. M. Kainerstorfer et al., “Principal component model of multispectral data for near real-time skin chromophore mapping,” J. Biomed. Opt., 15 (4), 046013 https://doi.org/10.1117/1.3470094 JBOPFO 1083-3668 (2010). Google Scholar

47. 

R. L. P van Veen et al., “Determination of visible near-IR absorption coefficients of mammalian fat using time- and spatially resolved diffuse reflectance and transmission spectroscopy,” J. Biomed. Opt., 10 (5), 054004 https://doi.org/10.1117/1.2085149 JBOPFO 1083-3668 (2005). Google Scholar

48. 

G. M. Hale and M. R. Querry, “Optical constants of water in the 200-nm to 200-μm wavelength region,” Appl. Opt., 12 555 –563 https://doi.org/10.1364/AO.12.000555 APOPAI 0003-6935 (1973). Google Scholar

Biography

Laura Lucía Sánchez-Ramos is a MS student in mechanical engineering at the Universidad Autónoma de San Luis Potosí. She received her BS degree from the Universidad Autónoma de San Luis Potosí. Her area of research is biomedical optics.

Beatriz Morales-Cruzado is a researcher at the Universidad Autónoma de San Luis Potosí. She received her MS degree and PhD in sciences with a specialty in optics from the National Institute of Astrophysics, Optics and Electronics in 2008 and 2012, respectively. Her research interests include biomedical optics, optical tweezers, and cavitation induced by laser. She has 22 articles in refereed journals and belongs to the National System of Researchers, level I.

Francisco Gerardo Pérez-Gutiérrez received his PhD in mechanical engineering from the University of California at Riverside. He is a researcher at the Universidad Autónoma de San Luis Potosí. His current research interests include biomedical optics, multiple optical trapping, and laser-matter interaction. He is the author of more than 25 journal papers and belongs to the National System of Researchers, level I.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Laura Lucía Sánchez-Ramos, Beatriz Morales-Cruzado, and Francisco Gerardo Pérez-Gutiérrez "Determination of tissue oxygen saturation by diffuse reflectance spectroscopy," Journal of Biomedical Optics 28(9), 095002 (28 September 2023). https://doi.org/10.1117/1.JBO.28.9.095002
Received: 27 February 2023; Accepted: 13 September 2023; Published: 28 September 2023